IGEA Cliniporator EPS02 User Manual page 29

Table of Contents

Advertisement

Cliniporator model EPS02 – User manual
2. bring the TAG placed on the
electrode packaging close to the
right part of the CLINIPORATOR,
identified by the symbol
hold it in this position until the
system recognises the electrode.
If the recognition is successful, a
window (Figure 31) provides the code
for the electrode recognised and the
operator is asked not to move the
packaging
so
"marking"
completed.
This process is mandatory according to
security obligations and allows the use
of the same electrode for a single
patient for 6 hours only.
To
mark
the
electrode
automatically configure the system for
use, press the confirmation button
without removing the packaging.
During the identification process, a
dialogue box (Figure 32) informs that
the operator should hold the TAG
present on the packaging close to the
antenna area
.
Interrupting the identification process can cause partial writing of information on the
electrode and therefore the electrode might become unusable. 
, and
can
be
and
to
WARNING!
Figure 30 – Electrode packaging positioning
Figure 31 – Electrode recognised
Figure 32 – Electrode marking
29