Electrode selection

5.6
After identifying all the electrodes needed for the treatment through the process described in
section 5.4, select the electrode to be used by pressing the indicator button placed to the right
and to the left of the electrode image (Figure 37 ), which could result enabled or not
depending on the number of electrodes loaded, and then press the electrode image. At the end
of the identification process, the last recognised electrode is the one displayed on the window,
since it is automatically selected.
The selected electrode is identified through its code displayed against a green background.
In the field
are shown messages that suggest the operation that the user can perform.


At the same time in the upper part of the screen (Figure 37 ) the treatment data associated to
the electrode are shown.
Number of delivered pulses.
Amplitude (in volt) of pulses.
Length (in microseconds) of pulses.
Repetition frequency (in Hz).
These parameters cannot be modified in the ECT treatment mode.
Pressing the forward button
32
Figure 37 –Electrode selection in "ECT" treatment
, the treatment screen is shown (section 5.9).
Cliniporator model EPS02 – User manual