Some electrode models must be connected to CLINIPORATOR using the HANDLE supplied
together with the CLINIPORATOR.
Accessory Type
Handle
IG0M910
Handle
IG0M915
The electrode for the treatment should be chosen according to the type of ECT or EGT
treatment, the size of the lesions to be treated or their position:
 the hexagonal arrayed needle electrodes (series EPS/EPSA) are used for the treatment
of big and deep lesions
 the parallel row arrayed needle electrodes (series EPS/EPSA) are used for the
treatment of small and deep lesions
 the plate electrode (series EPS), made up of two stainless steel plates, can be used for
the treatment of surface lesions without penetrating the skin
 the wearable finger electrodes (series NF/NFD) are used for the treatment of lesions in
small anatomical cavities.

Handle for Connecting the Electrodes

6.2
S o m e e l e c t r o d e m o d e l s m u s t b e c o n n e c t e d t o CLINIPORATOR using the Handle,
accessory supplied together with C L I N I P O R A T O R .
There are two handle models, the Blue and the Green one. They are both designed for multiple
use and different patients and also for sterilisation using the methods described in section 6.2.2.
The two models are distinguished by both the handle color, blue and green, and the
configuration of the electrical contacts at the end of the handle.
Each electrode can be connected to only one of the two supplied handle models according to
the instruction sheet of the specific electrode. Some electrode models do not require any handle
for the connection with C L I N I P O R A T O R .
All the instructions detailed related to the cleaning, sterilization, care and substitution of
the handle
are described in the illustrative sheet furnished in endowment with the
handle; read carefully the illustrative sheet of the handle before use.
6.2.1

Connecting/Disconnecting the handle to CLINIPORATOR

If the connection to the electrode requires the use of the handle, after having completed the
electrode recognition procedure described in section 5.4, connect the handle compatible with
the electrode to the proper front panel connector of C L I N I P O R A T O R ( reference 3 i n
Figure 3) . Reference to Figure 63:
48
Cliniporator model EPS02 – User manual
Description
Blue Handle
Re-sterilizable handle for needle electrodes, EPS
series, in linear configuration
Green Handle
Re-sterilizable handle for needle electrodes, EPS
series, in exagonal configuration.