Cliniporator model EPS02 – User manual
User interaction with C L I N I P O R A T O R is similar to that of a PC; the operator dialogues with
the device using the PC panel ( Figure 4 ) equipped with touch-screen display. Data are entered
by means of a virtual keyboard represented and controlled on the display, while the mouse
functions by moving the finger on the screen just like a mouse pointer.
The PC Panel is equipped with input/output ports located in the lower part. These connections
are used to connect the internal radio device (serial port) and the power unit found in the lower
part of the system (USB port). A further USB port allows connection with the external devices
necessary for data transfer, software update etc...
Reset button
During patient treatment:
 the input/output ports must only be used to connect devices compliant with the IEC
60601-1 standard;
 only devices that are powered by the port itself - and that, therefore, are not
connected to the power grid - must be connected to the USB ports, in addition to
devices compliant with standard IEC 60601-1
 the LAN network connection port must not be used to transfer data.
Figure 4 – PC Panel front side and layout of connectors in the lower part
WARNING