Cliniporator model EPS02 – User manual
In the case where the selftest shows an
electrode automatic recognition fault (see
section 4.7) and, consequently, only the
manual recognition procedure is allowed,
the notice in Figure 34 will appear by
pressing the "electrode selection" button
(Ref.  in Figure 28).
Pressing the confirmation button again
allows you to switch to the manual input
mask of the electrode safety code as
shown in Figure 37
Figure 34 – Manual electrode recognition notice
If the detection fails three times in a row, at
the next press of the confirmation button
the manual input of the electrode safety
code of Figure 37 is displayed.
By pressing the confirmation button again,
the manual electrode security code entry
screen is displayed as seen in .Figure 35.
First, select the electrode that you want to
use by clicking on the button
or
.
Figure 35 - ID code entry screen
Then, enter the 20-character code (ID CODE)
found on the sticker located on the electrode
bag
in
the
5
blank
boxes
using
the
alphanumeric keyboard.
Finally, confirm the settings by pressing the
ENTER button; the system checks the
coherence between the electrode selected
and the security code entered. If the values
entered are correct, the electrode code
window becomes green and the confirmation
button
is activated , which allows you to
move on to the "marking" phase. To cancel
the process without identifying the electrode,
press the button
.
Figure 36 – Electrode recognized