Cliniporator model EPS02 – User manual
Error message
Possible causes
ECG signal not
ECG signal not detected.
found
Attempt to use an electrode that is
Electrode code is
not approved by IGEA.
duplicated
More than six hours have passed
Electrode validity
since the electrode's last use on the
time elapsed
patient.
The
Electrode was
electrode between the recognition
changed
phase and the identification one.
Error
saving
treatment
data.
Error during data treatment saving.
Treatment has been
completed
correctly
but data will be lost
The USB device used is defective.
The recognition of the USB device is
Error while mounting
interrupted by a hardware / software
USB pen drive
defect.
Errors occurred during the data
Export completed
export process.
with errors
The password entered during login is
Incorrect password,
incorrect.
please retry!
Tag recognised as non valid.
Not a valid electrode
Electrode code not allowed in the
Not allowed in this
treatment mode chosen
mode
The system tried to perform an
Not in proper state
operation at a wrong time
Detected one or both pedals pressed
Pedal
pressed
during the self-test
detected
The ECG synchronizer is turned off.
The connection between the device
Please try to fix the
and the ECG synchronizer is not
ECG
signal
and
correct.
press the retry button
The
abnormally.
The RFID tag is not close enough to
the relevant antenna.
In the recognition phase the user
does not position the RFID tag near
the relevant antenna, or removes the
RFID
Timeout in electrode
recognition operation is completed.
recognition
The electrode RFID tag is damaged.
Hardware / software problem.
The PULSE pedal pressure has
Timeout on arming
been performed much later than the
pedal
ARM pedal pressure.
operator
has
changed
ECG
synchronizer
works
tag
before
the
electrode
What to do
To solve the problem, check ECG electrodes position
and connections.
Replace the electrode with an electrode approved by
IGEA.
Replace the electrode with a new one.
the
Repeat the recognition of the electrode by not
changing or moving the electrode between the
recognition phase and the marking phase..
Continue with the
treatment. If the problem persists,
contact the Customer Service of IGEA S.p.A.
Replace the USB device.
Restart the device and repeat the procedure. If the
problem persists, contact the Customer Service of
IGEA S.p.A.
Check if the USB pen drive is correctly inserted and if
it has enough free space.
drive, boot the device and retry the operation.
problem persists, contact the Customer Service of
IGEA S.p.A.
Provide the correct credentials
Repeat electrode selection. If the problem persists,
restart the device. If the problem persists, contact the
Customer Service of IGEA S.p.A.
The electrode chosen is not habilitated for EGT mode.
Select only electrodes habilitated for EGT mode.
If the system works, continue using it. In any case,
contact the Customer Service of IGEA S.p.A.
Turn off the device, check if pedals are properly
connected and not pressed and restart. If the problem
persists,
contact Customer Service of IGEA S.p.A.
Turn on the ECG synchronizer.
Check that the connection between Cliniporator and
the ECG synchronizer is correct (refer to the user
manual of the synchronizer).
Switch the synchronizer off and on again. If the
problem persists, contact the Customer Service of
IGEA S.p.A.
Bring the electrode Tag closer to the antenna area and
repeat the electrode recognition.
Bring the electrode Tag close to the antenna area and
repeat the electrode recognition by keeping the Tag in
position for the entire duration of the operation.
Recognize the electrode using the manual recognition
operation. Contact Customer Service of IGEA S.p.A.
Recognize the electrode using the manual recognition
operation. Contact Customer Service of IGEA S.p.A.
The ARM and PULSE pedals must be pressed
consequently within 10 seconds. If the problem
persists, Contact Customer Service of IGEA S.p.A.
Disconnect the
USB pen
If the
53