Cliniporator model EPS02 – User manual
Pressing the icon
activates the virtual
keyboard that allows you to enter explanatory
notes about the treatment performed.
Pressing the confirmation button
removes
it by saving the entered data, pressing the
button
cancels the entered data.
Figura 44- Virtual keyboard
Once started, the program waits until the operator selects the number of the nodule by pressing
the increase/decrease buttons
or charges the capacitors to start the treatment.
To charge the capacitors and accumulate energy press the charge button
, which is yellow at
the beginning .
The ongoing charging process is graphically represented by a "battery" shown in the button
that fills up progressively.
When the charging process is completed, the battery is completely full and the button turns
green (Figure 45
). At this stage, the treatment can be delivered by acting on the double pedal
for activation and pulse (Figure 2).
By pressing the activation pedal (white pedal in Figure 2) a 10 seconds reverse count starts
and - until complete delivery of the treatment - the buttons    in Figure 43 are replaced by
the "abort" button
in Figure 45. By pressing it the treatment is immediately interrupted; in this
case the system provides the message "Treatment aborted by user" and proceeds to save the
data of the treatment provided until the interruption.
Figure 45 - activation pedal pressed: wait pedal pulse
37