Non-Recognition Of The Electrode - IGEA Cliniporator EPS02 User Manual

Table of Contents

Advertisement

If the identification goes through, the system goes back to the "ELECTRODE SELECTION"
screen in Figure 37, where the recognised and identified electrode has already been entered in
the electrode list for use (Figure 37,
If during the identification process the system detects an error, a dialogue box appears
informing about the possible causes and the process is stopped. The following table shows the
most common error messages that might be displayed.
"Error in electrode recognition"
"Please check the tag position"
"Error selecting electrode"
"Error reading electrode"
"Error burning electrode"
"Electrode validity time elapsed"
"Electrode code is duplicated"
"Electrode has been changed"
It is possible to recognize and identify more than one electrode by repeating the above
mentioned process. Otherwise, start the treatment by pressing the "forward" button
bottom-right area which has now become active.

Non-recognition of the electrode

5.5
If the TAG on the electrode packaging is
damaged or the packaging is improperly
positioned, at the end of the automatic
electrode
recognition
message in Figure 33 is displayed .
By pressing the confirmation button the dialogue box is closed. Remember that, for this reason,
the electrode packaging must be correctly positioned as seen in Figure 30.
If this process fails three times in a row, when the confirmation button is pressed again, the
manual electrode security code entry screen appears as shown in Figure 35
30
).
The electrode TAG was not held close (maybe not
close enough) to the indicated position; this message
may also appear during the recognition phase.
An error occurred while trying to access the
electrode data.
Attempt to use an already used electrode for the
same patient after 6 hours from the first use.
The electrode has an invalid or duplicate code and
cannot be used for security reasons.
The electrode was replaced during the automatic
recognition phase.
process
the
error
Cliniporator model EPS02 – User manual
Figure 33 - Electrode recognition error