Power Supply Cord Replacement

8.6
To replace the power supply cord use a 4mm Allen key. This procedure is described below:
 unscrew the screw which secures the
cable clip to the rear panel of the
device
 open the cable clip and remove it from
the old cable
 apply it to the new cable, at a distance
of 23 cm from the VDE plug tip
 secure the cable clip to the rear panel
using the Allen screw, putting the star
washer between the cable clip and the
rear panel.
58
Cliniporator model EPS02 – User manual
Figure 65 – power supply cord