4 INSTALLATION AND START UP

4.1 System Location and Conditions of Use

As all electromedical devices, C L I N I P O R A T O R necessitates specific precautions in regards
to the electromagnetic compatibility (EMC) a n d must be installed and placed in a suitable
location which follows the conditions specified in Section 9.3 and 9.9.
Portable and mobile radio communication devices might influence the device operation.
C L I N I P O R A T O R must be used only in the operating room or in a medical clinic which meets
the requirements provided by the regulations in force.
The device is equipped with wheels which ensure an adequate movement. However, attention
must be paid during its handling, since the equipment cannot move over obstacles higher than
10 mm. In this case - using the front grip - the device has to be tilted in order to lift the wheels.
C L I N I P O R A T O R must be installed leaving at least a 10 cm free space at the back of the
device to allow ventilation. Do not use the equipment if the ventilation grids are blocked.
Connect C L I N I P O R A T O R exclusively to a grounded mains system.
C L I N I P O R A T O R must not be used in the presence of flammable mixtures.
C L I N I P O R A T O R is not protected against liquid penetration. Do not use any part of the
system as a surface support for containers with liquids.
Before beginning the treatment, always activate wheel locking levers and take the
appropriate precautions to avoid device accidental movement or handle cable yanking during
the treatment. Unexpected movements of the electrode could cause harm to the patient.

4.2 System Installation

C L I N I P O R A T O R must be installed only by specialized IGEA staff.
The parameters which allow the operators to access the system (username and password) are
entered during the installation procedure. In this phase the system checks also the correct
operation of the device.

4.3 Voltage Supply Selection

C L I N I P O R A T O R allows the selection between two
mains voltages: 115V and 230V. The voltage supply
selector is in the monobloc on the rear panel.
In order to select the voltage supply, carry out the
following steps:
1. make sure that the mains switch is in "O" position,
i.e. turned off,
2. make
sure
that
disconnected,
3. open the voltage selector cover by using a slotted
screwdriver,
16
WARNING!
the
power
supply
Cliniporator model EPS02 – User manual
cord
is
Figure 6 - Mains voltage monobloc