Minolta CA-100Plus Instruction Manual page 95

Crt color analyzer
Table of Contents

Advertisement

Command
*1
J
Sets calibration mode
when CA-100Plus is
set to analyzer (RBG)
display mode (avail-
able only with an op-
tional Analyzer Card)
R
For input of RGB emis-
G
sion characteristics for
B
CRT phosphors [available
W
only in analyzer (RBG)
display mode when using
optional Analyzer Card]
E
ENTER
K
Recalls standard color
data from memory
Z
Requests status infor-
m a t i o n r e g a r d i n g
present settings
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Function
"J0": Calibration mode off
"J1": Calibration mode on
"R": Sets R emission value
"B": Sets B emission value
"G": Sets G emission value
"W": Sets W (standard white value)
"E"
"K memory channel number "
Acceptable range: 00 to 10 (If op-
tional Card inserted, acceptable range
becomes 00 to 99.)
"Z"
Input format
92
Further information
• If the command "J0" (calibration
mode off) is input after RBGW data
has been set but before the com-
mand "E" (ENTER) is input, the
RBGW data just set will be deleted.
• If the command I, M, A, C, N, or P
is input, calibration mode will au-
tomatically be canceled.
• Corresponds to pressing
when CA-100Plus is in analyzer
mode.
• Will be accepted only when calibra-
tion mode has been set (by pressing
key or by inputting the com-
mand "J1").
• Correspond to pressing 4, 1, 0, and
7 respectively after
been pressed in analyzer mode.
• Sets present displayed data as stan-
dard color data in xyLv or T∆uvLv
display modes (or in analyzer mode
if calibration mode is off).
• Stores previously set RBGW val-
ues in memory when calibration
mode is switched on (if RBGW val-
ues have all been set correctly).
• Corresponds to pressing
• Standard color data is output for all
probe connector numbers selected
for data output.
• For format of output data, see page 97.
• The data output as a result of this
command must be accepted before
the next command is input.
• The following status information
will be output:
a. Display mode
b. SYNC mode
c. Memory channel number
d. ID label
e. Analog display range
f. Luminance units
g. Probe connector number for
which data is displayed
h. Probe connector number (or
numbers) for which data is out-
put
i. Calibration mode on or off
j. RBGW values set or not
• For format of output data, see page
94.
• The data output as a result of this
command must be accepted before
the next command is input.
key
key has
key.