When The Optional 4-Point Expansion Board Ca-B04 Is Used - Minolta CA-100Plus Instruction Manual

Crt color analyzer
Table of Contents

Advertisement

2. Installing the 4-Point Expansion Board CA-B04
Installing the optional 4-point expansion board CA-B04 in the instrument allows simultaneous measurement of the
colors or flicker at up to 5 points on the display's surface. Install the expansion board as shown below.
[Installation Method]
1
.
Remove the cover of the 4-point expansion board slot.
1 Set the POWER switch on the instrument to OFF.
2 Remove the two screws from the slot cover, and re-
move the cover.
2
.
Install the 4-point expansion board.
1 Place the 4-point expansion board along the right-
and left-side guides in the slot.
2 Push the board all the way and make sure that the
board is connected properly.
3 Secure the board with the two screws that were re-
moved previously.
• Repeatability of the measurement value becomes
worse when the fixation by the screw is incomplete.
• To remove the board, remove the two screws first,
then hold the grip of the board and pull it out. After
the board is removed, attach the cover to the slot.
<Notes on Installation>
When installing/removing the 4-point expansion board, al-
ways set the POWER switch to OFF and pull the AC power
cord from the AC outlet first.
Do not touch the connectors (gold plated parts) or ICs on the 4-point expansion board with your hands. If oil or
similar matter adheres to the connectors, wipe them with a soft, dry cloth.
<Connecting Measuring Probes>
The following two types of measuring probes can be connected.
• CA-P02 (cord length: 2 m)
• CA-P05 (cord length: 5 m)
A total of five probes can be connected. When connecting two
or more probes, always make sure that one of them is con-
nected to the probe connector [P1].
Connect necessary number of probes to the probe connectors
[P2] to [P5] on the 4-point expansion board. You do not have
to connect any probes to those connectors ([P2] to [P5]). Probes
can be connected to any connectors ([P2] to [P5]).
The connecting method for connectors [P] to [P5] is the
same as that for [P1]. (Refer to page 26.)
Notes when connecting probes: Probe connectors on the 4-point ex-
pansion board where no probe is connected must be capped.