<Error Messages in LCD Display Section>
"NO SYNC. SIGNAL" (when EXT mode is selected)
• Cause 1 : The vertical synchronizing signal used for the display is not connected to the terminal on the
instrument.
Action
: If EXT mode is selected, input the vertical synchronizing signal to the terminal on the instru-
ment properly.
• Cause 2 : The frequency of the vertical synchronizing signal used for the display is below 40 Hz or beyond
200 Hz.
Action
: Change SYNC mode to UNIV. mode and start measurement.

3. Selecting the Measurement Mode

The following measurement modes are available.
Measurement Mode
xyLv mode
T∆uvLv mode
G-reference
G Standard
Analyzer mode
R Standard
R-reference
u'v'Lv mode
XYZ mode
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
Used to display/output chromaticity coordinates xy and intensity Lv.
(The analog display section shows ∆x, ∆y and ∆Lv.)
Used to display/output T (correlated color temperature), ∆uv (color difference from
blackbody locus) and luminance Lv.
(The analog display section shows ∆x, ∆y and ∆Lv.)
Used to display meeasurement RBG emission intensities as a percentage of the RGB
emission intensities target color(W). Analog display shows measured ratios R/G and
B/G, and ∆G
Used to display meeasurement RBG emission intensities as a percentage of the RGB
emission intensities target color(W). Analog display shows measured ratios G/R and
B/R, and ∆R
Used to display/output u'v' chromaticity coordinates (CIE 1976 UCS chromaticity
diagram) and luminance Lv.
(The analog display section shows ∆x, ∆y and ∆Lv.)
Used to display/output tristimulus values X, Y and Z.
(The analog display section shows ∆x, ∆y and ∆Lv.)
40
... For other error messages, refer to page 107.