Minolta CA-100Plus Instruction Manual page 113

Crt color analyzer
Table of Contents

Advertisement

<Relationship Between Probe Serial No. and Error Message "E1">
If "E1" appears, the cause of the error can be located easily by checking the serial no. of the probe used to make
settings and the current probe serial no.
• Probe serial no. used for making settings : Displayed when the MR key is held down for two to four sec-
• Current probe serial no.
Probe Serial No.
Displayed during
Measurement
Both lines show "00000000".
00000000
00000000
Upper line shows "00000000".
00000000
16790160
Different probe nos.
21593001
16790160
Same probe nos.
16790160
16790160
Upper line shows "00000000".
Lower line shows "99999999".
00000000
99999999
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
onds. (The buzzer sounds once immediately after the MR key is
pressed. It will also sound two and four seconds later if the key is
kept held down. Thus, to display the probe serial no. used for
making settings, release the key after the buzzer has sounded
twice in total.)
: Displayed in the PROBE selection screen when the
When xyLv, T∆uvLv, u'v'Lv or XYZ
measurement mode is selected
Cause : User calibration has not been per-
formed or the target color has not been
set for the currently selected memory
channel since shipment from the fac-
tory.
Action : Perform user calibration or set the tar-
get color.
Cause : The currently used measuring probe
is different from the one that was used
to set the target color.
Action : 1 Set the target color using the mea-
suring probe that was used for user
calibration.
2 Perform user calibration again us-
ing the currently connected mea-
suring probe.
Cause : The currently used measuring probe
is different from the one that was used
to perform user calibration and set the
target color.
Action : 1 Perform measurement using the
measuring probe that was used to
perform user calibration and set the
target color.
2 Perform user calibration and set the
target color .using the currently
connected measuring probe.
Cause : The settings made to the selected
memory channel have been lost from
the instrument's memory.
As a result, the default (factory) cor-
rection factor and ID name will be
used instead.
Action : Set them again.
110
110
keys are pressed together.
Cause and Action for "E1"
When analyzer measurement mode is
Cause : Neither the RGB emission character-
Action : Set the RGB emission characteristic
Cause : The RGB emission characteristic for
Action : Set the RGB emission characteristic for the
Cause : The measuring probe used to input the
Action : 1 Set the target color using the mea-
Cause : The currently used measuring probe is dif-
Action : 1 Perform measurement using the mea-
Cause : The settings made to the selected
Action : Set them again.
selected
istic for display's analyzer mode nor
target color (W) have been set for the
currently selected memory channel
since shipment from the factory.
for the display's analyzer mode and
target color (W).
the display's analyzer mode has not
been input.
display's analyzer mode. If you want to
set a target color different from the one that
was set when the emission characteristic
was set, set a new target color.
RGB emission characteristic for
display's analyzer mode is different
from the one used to set the target color.
suring probe that was used to input
the emission characteristic for the
analyzer mode.
2 Input the emission characteristic for
the analyzer mode again using the
currently connected measuring
probe.
ferent from the one used to input the RGB
emission characteristic for display's ana-
lyzer mode and set the target color.
suring probe used to input the RGB
emission characteristic and set the
target color.
2 Input the emission characteristic and
set the target color using the currently
connected measuring probe.
memory channel have been lost
from the instrument's memory.
As a result, the default (factory) ID
name will be used instead.