2. Inputting the RGB Emission Characteristic for Analyzer Mode

The RGB emission characteristic for analyzer mode must be input to each memory channel. When it is input, the
target color (W) must also be set.
To adjust white balance, the values of the white-balanced white must be entered as the white color (W).
If the RGB emission characteristic for the display's analyzer mode is input to a memory channel for which the
target color has already been set, the previously set target color will be deleted. The target color to be used is the
same as that for xyLv, T∆uvLv, u'v'Lv and XYZ measurement modes.
The RGB emission characteristic for the display's analyzer mode must be input for each display type (model).
Characters of displays vary with the display type (model). Because of this, measured values differ even if the same
color is measured.
Thus, a different memory channel must be used for each display type (model) to input the RGB emission charac-
teristic for analyzer mode.
[Operating Procedure]
1
When the optional 4-point expansion board CA-B04 is used
Select the probe no. for which the RGB emission characteristic for the display's analyzer mode is to be input. The
RGB emission characteristic for the display's analyzer mode can be input independently for each probe connector
([P1] to [P5]) for each memory channel.
1 1 1 1 1
Press the
key.
The LCD display section will switch to the menu selec-
tion screen.
2 2 2 2 2
Press the
key to open the PROBE selec-
tion screen.
Each time the
key is pressed, the screen will switch
in the order PROBE → SYNC → ID Name input →
RANGE → Measurement Speed → Number of Digits →
Calibration Standard → RS232C Baud Rate → PROBE.
3 3 3 3 3
Press the
key to display the probe no. you
want to select.
Each time the
key is pressed, the probe no. switches
in the order [P1] ... .
4 4 4 4 4
Press the
key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automatically when the POWER switch is set to ON. If you
want to change this setting, refer to page 29.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
3
8
2
4
56
59
4,5,6,7
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
PROBE selection screen
SELECT : PROBE
P1 35881112
SELECT : PROBE
P3 35881113
Probe no.
Press the
key
until the desired
probe no. appears.