Measurement/Quick Guide

Before starting measurement, perform necessary preparations as explained in the Installation/Connection section (page 23).
<Zero Calibration> Page 34
1. Check that the POWER switch is set to ON.
2. Set the Pointing ring of the measuring probe to the 0-CAL
position.
When the optional 4-point expansion board CA-B04 is used
Set the Pointing ring of every measuring probe to the 0-
CAL position.
3. Press the 0-CAL key.
The digital and analog display sections will light up and
measurement starts.
<Selecting Measurement Speed> Page 36
1.Press the
2. Press the
screen.
RS232C Baud Rate→Calibration Standard→Number of Digits→Measurement Speed
3. Press the
AUTO→SLOW→FAST
4. Press the
<Selecting SYNC Mode> Page 38
1. Press the
2. Press the
3. Press the
EXT→UNIV→INT→NTSC→PAL
4. Press the
* To use EXT mode, the vertical synchronizing signal used for the
display must be input to the instrument. (Page 28)
<Selecting Measurement Mode> Page 40
1. Press the MODE key until the desired measurement
mode appears.
xyLv→T∆uvLv→Analyzer (G standard)
XY←FLIC←u'v'Lv←Analyzer (R standard)
To the next page
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
key to display the menu selection screen.
key to open the measurement speed selection
PROBE→SYNC→ID Name input→RANGE
key to display the desired measurement speed.
key to confirm the selection.
key to display the menu selection screen.
key to open the SYNC selection screen.
key until the desired SYNC mode appears.
key to confirm the selection.
1
POWER switch
3
0-CAL key
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
Measurement speed selection screen
SELECT : M-SPD
AUTO
SELECT : M-SPD
SLOW
SELECT : M-SPD
FAST
"F" is displayed when the [FAST] was selected.
CH00 EXT
[
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
SYNC selection screen
SELECT : SYNC.
EXT
SELECT : SYNC.
UNIV
Measurement mode
1
Mode key
115
2 Pointing ring
Press the
key until the
desired
measurement
speed
appears.
Fd P1
]
Press the
key until the
desired SYNC
mode appears.