Turning the Power ON/OFF

1. Turning the Power ON ( | )/OFF ( )
Before setting the POWER switch to ON, prepare the following.
1
.
Connect a measuring probe to the probe connector [P1]. (Page 26)
• To synchronize measurement with the ... 1 Input the vertical synchronizing signal that is used for the display.
display's vertical synchronizing signal
(EXT is selected as the SYNC mode)
• To perform measurement........................ 1 Install the 4-point expansion board CA-B04 (option) in the in
with two or more measuring probes
• To communicate with the PC ................. 1 Connect the instrument's RS-232C connector to the PC. (Page 84)
via RS-232C ........................................... 2 From the menu selection screen, select the desired RS232C
• To communicate with the PC via USB ... 1 Connect the instrument's USB connector to the PC. (Page 86)
2
. Connect the AC power cord to an AC outlet. (Page 28)
[Turning the Power ON ( | )]
Set the POWER switch to ON.
If the instrument is connected to external equipment,
set the instrument's POWER switch to ON first, then
turn ON the power to the external equipment.
[Turning the Power OFF ( )]
If the instrument is connected to external equipment, turn OFF the power to the external equipment first,
then set the instrument's POWER switch to OFF.
<Error Messages in LCD Display Section>
"SET MAIN PROBE" (After the POWER switch is set to ON)
• Cause 1 : The measuring probe is not connected to the probe con-
nector [P1] properly.
• Action 1: Set the POWER switch to OFF, then connect the measuring probe to the probe connector [P1]
properly. (Before connecting/disconnecting the measuring probe, make sure that the POWER
switch is set to OFF.)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
(Page 28)
strument. (Page 27)
2 Connect necessary number of probes to the probe connec-
tors [P2] to [P5]. (Pages 26 and 27)
baud rate. (Page 85)
CA-200 MODE
PROBE [P1]
NO.XXXXXXXX
DARKEN PROBE
PUSH 0-CAL KEY
... For other error messages, refer to page 107.
29
29
Either "CA-200
MODE" or "CA-100
COMPATIBLE
MODE?" is displayed
(see page 13).
Probe serial no.
SET MAIN PROBE