Minolta CA-100Plus Instruction Manual page 58

Crt color analyzer
Table of Contents

Advertisement

[Preparation]
1
.
Press the MODE key to select xyLv measurement mode.
2
.
Press the MEMORY CH
channel where you want to perform user calibration ap-
pears.
A memory channel other than CH00 must be selected.
3
.
Place the measuring probe against the display and take
measurement.
Set the display so that it can display four colors (RGBW) whose xyLv val-
ues are known.
4
.
Press the
key.
The LCD display section will switch to the user calibration input screen.
5
.
Enter the emission characteristic of R and calibration val-
ues (x, y, Lv).
1 Place the measuring probe against the display, which is now emitting
monochrome light of R.
Currently measured values will be displayed.
2 While the probe is placed against the display, press the HOLD key.
The measured values will be hold and the HOLD lamp lights up.
3 Press the
key.
The LCD display section will switch to the R calibration value input
screen.
4 Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for
white calibration (see step 7 in "Performing White Calibration" on page
53).
5 Press the
key.
The LCD display section will return to the user calibration input screen,
with the "*" mark displayed on the left of "R".
6 Press the HOLD key to resume measurement.
6
.
Enter the emission characteristic of G and calibration values (x, y, Lv).
1 Place the measuring probe against the display, which is now emitting monochrome light of G.
Currently measured values will be displayed.
2 While the probe is placed against the display, press the HOLD key.
The measured values will be hold and the HOLD lamp lights up.
3 Press
key.
The LCD display section will switch to the G calibration value input
screen.
4 Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for white
calibration (see step 7 in "Performing White Calibration" on page 53).
5 Press the
key.
The LCD display section will return to the user calibration input screen,
with the "*" mark displayed on the left of "G".
6 Press the HOLD key to resume measurement.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
and
keys until the memory
55
CH01 EXT
Ad P1
[
]
CH01
U-CAL
P1
W
R
G
B
CH01 x
y
Lv P1
0000 0000
0.00
_
CH01
U-CAL
P1
W *R
G
B
"*" mark is displayed.
CH01
U-CAL
P1
W *R
G
B
CH01 x
y
Lv P1
0000 0000
0.00
_
CH01
U-CAL
P1 W *R *G
B
"*" mark is displayed.