Measurement; Performing Measurement - Minolta CA-100Plus Instruction Manual

Crt color analyzer
Table of Contents

Advertisement

Measurement

Before starting measurement, perform the following.
Installation/Connection section (page 23)
Measurement Preparation section (page 33)
Settings section (page 45)

1. Performing Measurement

[Measuring Method]
1
.
Press the MEMORY CH
the memory channel for which user calibration has been
performed (page 51), the RGB emission characteristic for
analyzer mode that has been input (page 59) and the tar-
get color that has been set/changed (page 62).
2
.
Place the measuring probe against the display and per-
form measurement.
The measurement results will be displayed in the digital and analog dis-
play sections in the selected measurement mode.
<About Low-Intensity Warning>
The digital and analog display sections are blinking.
• The measured Lv (intensity) is below 0.1 cd/m
or equivalent.
<Notes on Measurement>
Since the intensity of the display will be unstable for a while immediately after the display is turned ON, the
measured values must be read after they have stabilized.
Static electricity on the display's screen surface must be removed as much as possible.
Perform zero calibration if the ambient temperature has changed.
When measuring a display of intensity of 0.1cd/m
approximately every hour.
Make sure that the measuring probe is placed straight against the display. If it is tilted or moved, accurate
measurement cannot be performed.
Take care not to let the measuring probe be exposed to excessive impact. In addition, do not pull or bend the
cord excessively or exert excessive force on it. Failure to observe these cautions may result in breakdown or
wire-breakage.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Install the instrument, connect the power cable, and turn ON the power.
Perform preparations (instrument setting, zero calibration) that are
required prior to measurement.
Set up the instrument according to the setting method.
This is not necessary if the instrument has already been set up or if you
are going to perform measurement using Minolta's calibration standard
and are not going to use the analog display function
and
keys. keys to select
2
for white calibration with Minolta's calibration standard
2
or less for a long period of time, perform zero calibration