From the Measurement Preparation section
When performing measurement using Minolta calibration standard
<Selecting the Calibration Standard> Page 50
1. Press the
key to display the menu selection screen.
2. Press the
key to open the calibration stan-
dard selection screen.
3. Press the
key until the desired calibration
standard appears.
6500k ←→ 9300k
4. Press the
key to confirm the selection.
* The selected calibration standard will be set for CH00
as well as for all the memory channels that have not
been user-calibrated.
*1
<Setting/Changing the Target Color> Page 62
1. Setting/Changing the Target Color by Measurement Page 63
This method can be used for any memory channels.
4
3,5
1
HOLD lamp
When the optional 4-point expansion board CA-B04 is used
1 Press the
key to display the menu se-
lection screen.
2 Press the
key to open the PROBE se-
lection screen.
3 Press the
key until the desired probe
connector no. appears.
4 Press the
key to confirm the selection.
Memory channel
CH01 EXT
[
1. Press the CH
and
desired memory channel.
2. Place the measuring probe against the display and
take measurement.
3. Press the HOLD key.
The HOLD lamp will light up.
4. Press the
key.
5. Press the HOLD key. Measurement will start.
The HOLD lamp will light off.
To set an ID name
To use the analog display function : *3 <Setting the Analog Display Range>
To the Measurement section
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2
Probe connector no.
Ad P1
]
keys to select the
: *2 <Setting an ID Name> Page 119
Page 75
117
2. Setting/Changing the Target Color by Entering Values
Page 65
This method can be used for memory channel CH00 only.
2
1
3
4
When the optional 4-point expansion board CA-B04 is used
1 Press the
key to display the menu se-
lection screen.
2 Press the
key to open the PROBE se-
lection screen.
3 Press the
key until the desired probe
connector no. appears.
4 Press the
key to confirm the selection.
Probe connector no.
Memory channel
CH00 EXT
[
1. Press the CH
and
2. Press the
key.
CH00 x
y
3127 3293 160.0
_
Cursor
3. Enter the target color (x, y, Lv).
Ten-key (
to
,
Key ...................................... The cursor moves in
4. Press the
key.
Page 119
Ad P1
]
keys to select CH00.
Lv P1
) ... Used to enter values.
the order
x → y → Lv → x.