2. Setting/changing the target color by entering values

This method can be used for memory channel CH00 only.
[Operating Procedure]
When the optional 4-point expansion board CA-B04 is used
Select the probe no. to which you want to set the target color. The target color can be set independently for each
probe connector ([P1] to [P5]) for each memory channel.
1 1 1 1 1
Press the
key.
The LCD display section will switch to the menu selec-
tion screen.
2 2 2 2 2
Press the
key to open the PROBE selec-
tion screen.
Each time the
key is pressed, the screen will switch
in the order PROBE → SYNC → ID Name input →
RANGE → Measurement Speed → Number of Digits →
Calibration Standard → RS232C Baud Rate → PROBE.
3 3 3 3 3
Press the
key to display the probe no.
you want to select.
key is pressed, the probe no. switches in the order [P1] ... .
Each time the
4 4 4 4 4
Press the
key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automatically when the POWER switch is set to ON. If you
want to change this setting, refer to page 29.
1
.
Press the MEMORY CH
until the memory channel CH00 appears.
2
.
Press the
key.
In the LCD display section, the current target color val-
ues are displayed.
<Error Messages in LCD Display Section>
"E3" (after the
key is pressed)
• Cause : An attempt was made to set Incorrect target
color values.
Incorrect calibration values mean the following.
1 One of x, y and Lv is "0".
2 1– x– y
3 Values which are beyond the instrument's calculation capability or contradicting values.
• Action : Enter correct values and then press the
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
and
keys
0
65
2
4
1
3
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
PROBE selection screen
SELECT : PROBE
P1 35881112
SELECT : PROBE
P3 35881113
Probe no.
Memory channel
CH00 EXT
Ad P1
[
CH00 x
y
Lv P1
3127 3293 160.0
_
... For other error messages, refer to page 101.
E3
x y Lv P1
0000 3293 160.0
_
key.
Press the
key
until the desired
probe no. appears.
]

Advertisement