M&C EC-30C Instruction Manual page 31

Ultra low gas cooler
Table of Contents

Advertisement

Components
Alarm type
Pre-cooling unit
Temperature
Temperature
Deep-cooling
unit I
Peltier-
element(s)
0 - 20 V
Module 1
Temperature
Deep-cooling
unit II
Peltier-
element(s)
0 - 20 V
Module 2
EC-30C | 1.00.02
+49 2102 935 888
Symbol Code
Description
Temperature
01
too high
Temperature
02
too low
Cooling
10
temperature
too high
Heating
11
temperature
too high
12
Loss of power
13
defect
14
defect
Cooling
20
temperature
too high
Heating
21
temperature
too high
22
Loss of power
23
defect
24
defect
www.mc-techgroup.com
Comments
T1 ≥ +5 °C (41 °F)
T1 ≤ -1 °C (30.2 °F)
T2 ≥ T2
+3 °C (37.4
set
°F)
T2 ≥ +30 °C (86 °F)
U
≥ 19.7 V
PE1
No cooling despite
the FET, relay etc. OK
U
≤ U
-1 V
OUT
OUT, set
or
U
≥ U
+1 V
OUT
OUT, set
T3 ≥ T3
+3 °C (37.4
set
°F)
T3 ≥ +30 °C (86 °F)
U
≥ 19.7 V
PE2
No cooling despite
the FET, relay etc. OK
U
≤ U
-1 V
OUT
OUT, set
or
U
≥ U
+1 V
OUT
OUT, set
30

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EC-30C and is the answer not in the manual?

Table of Contents