Summary of Contents for M&C LC Series

  • Page 1 Series LC, LG, LT Gas and Liquid Cooler LGC-1(S), LGT-2 LC-1(S), LTC-1 Instruction Manual Version 1.00.01 Gas Cooler LGC-1(S) Liquid Cooler LC-1(S)
  • Page 2 Dear customer, Thank you for buying our product. In this instruction manual you will find all necessary information about this M&C product. The information in the instruction manual is fast and easy to find, so you can start using your M&C product right after you have read the manual.
  • Page 3: Table Of Contents

    Table of Contents General Information ............................. 4 Declaration of conformity ........................... 4 Safety instructions ..............................5 Warranty................................. 5 Used terms and signal indications ........................6 Introduction ................................8 Serial number ......................................... 8 Description ................................8 Technical Data ..............................10 Gas Cooler ..........................................10 Liquid Cooler ........................................
  • Page 4: General Information

    Head Office M&C TechGroup Germany GmbH  Rehhecke 79  40885 Ratingen  Germany Telephone: 02102 / 935 - 0 Fax: 02102 / 935 - 111 E - mail: info@mc-techgroup.com www.mc-techgroup.com General Information The product described in this operating manual has been examined before delivery and left our works in perfect condition related to safety regulations.
  • Page 5: Safety Instructions

    Safety instructions Follow these basic safety procedures when mounting, commissioning or operating this equipment: • These instruction manual points out the basic dangers when handling the cooler. Further system-related hazards may occur in individual cases, which the operator must evaluate according to his hazard assessment - and take appropriate measures.
  • Page 6: Used Terms And Signal Indications

    Used terms and signal indications The ‘Danger’ warning sign indicates that death, serious injury and/or significant material damage will be the consequence, if the appropriate precautions should not be taken. Danger The ‘Warning’ warning sign indicates that death, serious injury or damage to property may occur if the relevant precautionary measures are not observed.
  • Page 7 Wear safety glasses! Protect your eyes while working with chemicals or sharp objects. Wear safety glasses to avoid getting something in your eyes. Wear protective clothes! Working with chemicals, sharp objects or extremely high temperatures requires wearing protective clothes. www.mc-techgroup.com LC, LG, LT | 1.00.01...
  • Page 8: Introduction

    Introduction The M&C gas coolers LGC-1 (with cooling coil for sample gas), LGC-1S (with longer cooling coil for sample gas) and LGT-2 (with tube bundle for sample gas) are used in analytical technology to lower the dew point of humid sample gases.
  • Page 9: Figure 2 Liquid Cooler Lc-1 With Cooling Coil For The Liquid Sample

    The M&C liquid coolers LC-1 and LC-1S are designed with a cooling coil fully welded in the jacket tube. The LC-1S version is equipped with a longer cooling coil. Coolant IN Dial thermometer Figure 2 Liquid cooler LC-1 with cooling coil for the liquid sample A dial thermometer is positioned in the coolant inlet of the LC-1 and LC-1S liquid coolers (coolant IN).
  • Page 10: Technical Data

    Technical Data 8.1 Gas Cooler Gas Cooler LGC-1 LGC-1S LGT-2 Part No. 04K1000 04K1500 04K4000 Max. cooling capacity approx. 900 kJ/h 1600 kJ/h 3600/6100 kJ/h Dimension cooling coil 1 x 4/6 mm 1 x 4/6 mm Cooling coil length 3.5 m [≈ 11.5 ft] 6 m [≈...
  • Page 11: Liquid Cooler

    8.2 Liquid Cooler Liquid cooler LC-1 LC-1S LTC-1 Part No. 04K2000 04K2500 04K3000 Max. cooling capacity approx. 900 kJ/h 1600 kJ/h 3600/6100 kJ/h Dimension cooling coil 1 x 4/6 mm 1 x 4/6 mm Cooling coil length 3.5 m [≈ 11.5 ft] 6 m [≈...
  • Page 12: Gas Cooler Dimensions

    8.3 Gas Cooler Dimensions A Sample gas IN B Sample gas OUT C Condensate OUT D Coolant IN Coolant OUT Figure 3 Dimensions: gas cooler LGC-1(S) A Sample gas IN B Sample gas OUT C Condensate OUT D Coolant IN Coolant OUT Figure 4 Dimensions: gas cooler LGT-2...
  • Page 13: Liquid Cooler Dimensions

    8.4 Liquid Cooler Dimensions A Liquid sample IN B Liquid sample OUT D Coolant IN Coolant OUT Figure 5 Dimensions: liquid cooler LC-1(S) A Liquid sample IN B Liquid sample OUT D Coolant IN Coolant OUT Figure 6 Dimensions: liquid cooler LTC-1 www.mc-techgroup.com LC, LG, LT | 1.00.01...
  • Page 14: Reception And Storage

    Reception and Storage The cooler must be stored in a weather protected frost-free area! During transport and storage, the cooler should always be positioned and stored according to the position indicated on the packaging. Note Preparation for Installation Select the optimum installation site in accordance with the generally applicable guidelines or consult the responsible authorities.
  • Page 15: Commissioning

    Commissioning The work described in this chapter needs to be performed by qualified personnel. The following minimum knowledge is required for the work: • Instructed person in the process engineering field. • Detailed knowledge of the operating instructions and the applicable safety regulations.
  • Page 16: Closing Down

    Closing Down Escape of potentially harmful gas from open process connections after disassembly possible. The relevant safety regulations must be observed for the media to be Danger conveyed. Before opening flush the lines with a suitable inert gas. Aggressive condensate possible! Chemical burns caused by aggressive media possible! For general electrical and mechanical work on the cooler, wear personal protective equipment (PPE) in accordance with the risk assessment.
  • Page 17: Maintenance

    Maintenance The safety instructions specific to the plant and process are to be consulted prior to any maintenance work! The work described in this chapter needs to be performed by qualified personnel. The following minimum knowledge is required for the work: •...
  • Page 18: Proper Disposal Of The Device

    Proper disposal of the device At the end of the life cycle of our products, it is important to take care of the appropriate disposal of obsolete electrical and non-electrical devices. To help protect our environment, please follow the rules and regulations of your country regarding recycling and waste management.