Routine Maintenance; Model 300M Maintenance Schedule; Table 11-1: Maintenance - Teledyne 300M Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

11 ROUTINE MAINTENANCE

The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.

11.1 Model 300M Maintenance Schedule

The following are the recommended periodic maintenance items for the Teledyne API Model
300M CO Analyzer:
Table 11-1: Maintenance Schedule
Item
JAN
Particulate
filter element
Pump
diaphragms
Sample flow
Dark Cal
Pneumatic
lines
Leak Check
Clean optical
bench
NOTE
Date Instrument Received _____________
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
11-1
Recommended Action
OCT
NOV
DEC
Replace weekly or as
needed
Replace every 12
months
Check for proper flow
(800cc/min ± 20%)
annually
Perform annually as
needed, or after
adjusting R7 on Sync-
Demod board.
Examine and clean as
necessary
Leak Check after
maintenance or repair
and at least annually
As needed

Advertisement