Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 10-1: Warning Messages
Warning Message
SOURCE WARNING
SYNC ERROR
BENCH HEAT
SHUTDOWN
SAMPLE PRESSURE
WARNING
SAMPLE FLOW
WARNING
BOX TEMP
WARNING
SAMPLE TEMP
WARNING
CANNOT DYN ZERO
CANNOT DYN SPAN
V/F NOT INSTALLED
SYSTEM RESET
RAM INITIALIZED
Meaning
The CO REF value is greater
than 5000 mV or less than
2500 mV
No modulation is present on the
output of the IR detector.
Temperature control of the
Optical bench cannot be
maintained at its 48°C set point
The Sample Pressure is less
than 15"Hg or is greater than
35"Hg
The sample flow is less than
500 cc/min or greater than 1200
cc/min.
The inside chassis temp is less
than 10°C or is greater than
50°C
The Sample Temperature is less
than 10°C or is greater than
50°C
An offset of more than ±5 PPM
would be required to Zero
adjust the analyzer
A slope of less than 0.5 or
greater than 2.0 would be
required to Span adjust the
analyzer
The CPU is unable to
communicate with the V/F
Board
A power Off-On cycle has
occurred
Dynamic memory has been re-
initialized in response to the
installation of a new PROM or
memory chip
10-4
Corrective Action
Check and adjust the Sync Demodulator
and optical alignment as described in
Section 10.6.6 and 10.6.5
Check IR source, IR detector/pre-amp,
and Opto interrupter
Check Optical Bench heater and
thermistor
Check for pressure transducer problems
as described in Section 10.6.5 and
Section 10.6.1
Check for pneumatic system problems as
described in Section 10.6.1. Check for
flow transducer problems as described in
Section 10.6.5
See Section 10.6.2
See Section 10.6.2
See Section 10.5.5
See Section 10.5.6
Check and re-seat CPU and V/F board.
See Section 10.6.3
None required
None required