Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

1.5 Installation and Overview

The Model 300M is shipped with the following standard equipment:
1. Power cord
2. Operation manual
To avoid personal injury, always use two persons to
Upon receiving the Model 300M please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise shipper first,
then Teledyne API.)
2. Before operation it is necessary to remove the shipping hold down screws. Remove the
instrument cover, then refer to Figure 1-2 for screw location.
3. When installing the Model 300M, allow a minimum of 4 inches (10 cm) of clearance at the
back of the instrument and 1 inch (2.5 cm) of clearance on each side for proper ventilation.
Also, be sure that the clearance below the chassis is unobstructed by at least the height of the
instrument feet.
4. Connect sample inlet line to the sample port on rear panel.
See Figure 1-3 for rear panel pneumatic connections. Sample gas
should only come into contact with PTFE, glass or stainless steel.
Connect the exhaust fitting on the rear panel (see Figure 1-4) to a
5. Connect a recording device to the terminal strip connections on the rear panel (See
Figure 1-3).
6. Connect the power cord to an appropriate power outlet (see the serial number tag for correct
voltage and frequency).
CAUTION
lift and carry the Model 300M.
NOTE
Leak check all fittings with soap solution.
CAUTION
suitable vent outside the analyzer area.
1-6