Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 1-2: Final Test and Calibration Values
Test Values
CO MEAS
CO REF
MR RATIO
SAMPLE PRESS
VACUUM PRESS
SAMPLE FLOW
BENCH TEMP
DC POWER
SUPPLY
SAMPLE TEMP
BOX TEMP
WHEEL TEMP
Dark Ref
Dark Meas
Electrical test
CO Span Conc
CO Slope
CO Offset
Noise at Zero (p-p)
Noise at Span (p-p)
Factory Installed Options
Factory Installed Options
Power Voltage/Frequency
Rack Mount, w/ Slides
Internal Zero Span Valves
Analog Voltage Range
PROM #
Date
Observed
Units
Value
mV
mV
in-Hg
in-Hg
CC/MIN
°C
mV
°C
°C
°C
Diagnostic
mV
mV
PPM
Span and CAL Values
PPM
PPM
PPM
Option Installed
Option Installed
0 - V
Serial #
Technician
1-16
Nominal Range
2500 - 4700
2500 - 4700
1.1 – 1.22
27 - 30
10 ± 5
700 - 900
48 ± 1
2500 ± 200
45-50
8- 40
68 ± 2
125 ± 50
125 ± 50
400 ± 4
400
1.0 ± 0.3
-0.1 to +0.1
0.5
<1.0% of reading