Table 5-2: I/O Signals - Teledyne 300M Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 5-2: I/O Signals (Continued)
#
Signal
8
WHEEL_HTR
9
DARK_CAL
10
ELEC_TEST
11
BROWNOUT_RESET
12
ST_CONC_ALARM_1
13
ST_CONC_ALARM_2
14
ST_FLOW_ALARM
15
ST_TEMP_ALARM
16
ST_DIAG_MODE
17
ST_POWER_OK
18
ST_PRESS_ALARM
(table continued)
Control
Description
YES
Shows the status of the filter wheel heater. This has
the same function as the LED in the power supply
module.
YES
Turns off the detector input to the Sync/Demod board
for electronics calibration.
YES
Activates the Electric Test diagnostic circuitry
YES
Resets the DISP_BROWNOUT circuitry described
above
YES
Status Bit - gas alarm 1
Logic high = M300M in alarm 1 mode
Logic low = Not in alarm 1 mode
YES
Status Bit - gas alarm 2
Logic high = M300M in gas alarm 2
Logic low = Not in alarm 2 mode
YES
Status Bit - Flow alarm
Logic High = Sample flow out of spec
Logic Low = Flows within spec
YES
Status Bit - Temperature alarm
Logic High = One or more temps out of spec
Logic Low = Temps within spec
YES
Status Bit - In Diagnostic mode
Logic High = M300M in Diagnostic mode
Logic Low = Not in Diag mode
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
YES
Status Bit - Flow alarm
Logic High = Sample pressure out of spec
Logic Low = pressure within spec
5-4