1.4 Specifications

Ranges
Zero Noise
Span Noise
Lower Detectable Limit
Zero Drift (24 hours)
Zero Drift (7 days)
Span Drift (7 days)
Linearity
Rise/Fall Time (95%)
Sample Flow Rate
Temperature Range
Humidity Range
Dust
Temp Coefficient
Voltage Coefficient
Dimensions HxWxD
Weight
Power
Power, CE
Recorder Outputs
Status Output
Alarm output
* at constant temperature and voltage
User selectable to any full-scale range from 0 - 30 PPM up to 0 - 3,000 PPM
< 0.1 PPM (rms)
< 0.5% of reading (rms) or 0.1 PPM (rms)
0.2 PPM
* <0.5 PPM
* <1.0 PPM
* 1% of reading
1% of Full Scale
<60 sec to 95%
800cc/min. ± 10%
5-40°C
0-95%RH, non-condensing
Sample must be dust free
< 0.5 PPM per °C
< 0.5 PPM per Volt
7" x 17" x 25" (178 mm x 432 mm
50 lbs (22.7 kg)
100 50/60 Hz, 115 60 Hz, 220V 50/60 Hz., 240 V 50 Hz. 250 Watts
230 V/50 Hz, 250 Watts
± 100 mV, ± 1 V, ±5 V, ±10 V (Bi-Polar)
9 status outputs from opt-isolators
3 opto-isolators to drive TTL relays
1-5
x 635 mm)