Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 5-2: I/O Signals (Continued)
#
Signal
19
ST_HIGH_RANGE
20
ST_SYSTEM_OK
21
ST_BENCH_ALARM
22
ST_SOURCE_ALAR
M
23
ST_WHEEL_ALARM
24
CO_MEASURE
25
CO_REFERENCE
26
SAMPLE_PRESURE
27
VACUUM_PRESSUR
E
28
SAMPLE_TEMP
29
BENCH_TEMP
30
WHEEL_TEMP
31
BOX_TEMP
(table continued)
Control
Description
YES
Status Bit - Autorange High Range
Logic High = M300M in high range
Logic Low = M300M in low range
YES
Status Bit - System OK
Logic High = No instrument warning present
Logic Low = 1 or more alarm present
YES
Status Bit - Bench Temperature Alarm
Logic High = Bench Temp out of spec
Logic Low = Bench Temp in spec
YES
Status Bit - IR Source Alarm
Logic High = IR Source output too low
Logic Low = IR Source output normal
YES
Status Bit - Wheel Temperature Alarm
Logic High = Wheel Temp out of spec
Logic Low = Wheel Temp in spec
NO
IR detector reading during measure phase. Typically
2500 - 4500 mV.
NO
IR detector reading during reference phase. Typically
2500 - 4500 mV.
NO
Sample pressure in mV. Typical sea level value =
4300 mV for 29.9" HG-A.
NO
Sample flow in mV
NO
Sample temp in mV
NO
Optical Bench temp. typically 2270 mV for 48°C
NO
Filter Wheel temp. typically 4770 mV for 68°C
NO
Internal analyzer temp. in mV
5-5