Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
After changing the variable's value, the CPU will respond with:
V VARIABLE=VALUE WARNLO WARNHI [DATALO-DATAHI] <CRLF>
which should reflect the new value. The values in square brackets are not required for all
variables. If needed, the values are included on the command line, separated by spaces. For
example, to change the instrument ID, the host computer would issue a command like this:
V MACHINE_ID=1234<CRLF>
and the CPU should respond with:
V DDD:HH:MM IIII MACHINE_ID = 1234 (0-9999)<CRLF>
Table 7-9 lists the variable names that are variable through the RS-232 interface and their
corresponding button sequences.
Table 7-9: RS-232 Variable Names
Variable Name
MACHINE_ID
BAUD_RATE
CURR_TIME
CURR_DATE
Button Sequence
SETUP-MORE-COMM-ID
SETUP-MORE-COMM-BAUD
SETUP-CLK-TIME
SETUP-CLK-DATE
7-12
Legal Values
0000 - 9999
300, 1200, 2400, 4800, 9600, 19.2
00:00 - 23:59
01/01/00 - 12/31/99