Troubleshooting; Overview - Teledyne 300M Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

10 TROUBLESHOOTING

The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.

10.1 Overview

The Model 300M has been designed to rapidly detect possible problems and allow their quick
evaluation and repair. During operation, the analyzer continuously performs self-check
diagnostics and provides the ability to monitor the key operating parameters of the instrument
without disturbing monitoring operations. These capabilities will usually allow the quick
isolation and resolution of a problem.
A systematic approach to troubleshooting will generally consist of the following four steps,
performed in order:
1. Confirm the proper operation of fundamental instrument sub-systems (Power Supplies, CPU,
Display).
2. Note any warning messages and take corrective action as required.
3. Examine the values of all TEST functions and compare to factory values. Note any major
deviations from the factory values and take correction action as required.
4. Address any dynamic (sample related) problems.
The following sections provide a guide for performing each of these steps. Figure 1-2 in this
manual shows the general layout of components and sub-assemblies in the analyzer and can be
referenced in performing the checks described in the following sections.
NOTE
10-1

Advertisement

Table of Contents
loading

Table of Contents