Rs-232; Pneumatic System; Table 1-1: Status Outputs - Teledyne 300M Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 1-1: Status Outputs

Pin Pair
Output#
(Low, High)
1
2
3
4
5
6
7
8
9
10
11
12

1.6.4 RS-232

The RS-232 connection is a male, 9-pin D-sub connector at the location shown in Figure 1-3.
See also Appendix A for additional information.

1.6.5 Pneumatic System

The Model 300M is equipped with a vacuum pump capable of pulling 800 cc/min across a flow
restrictor. This allows a smooth, stable flow of sample through the analyzer.
Sample enters the analyzer through a particulate filter element (47 mm diameter) mounted
immediately behind the front panel. The sample then enters directly into the sample cell. Please
see Figure 1-1 for a flow diagram and Figure 1-4 for pneumatic connections.
Definition
1,2
ALARM 1
3,4
ALARM 2
5,6
SYSTEM OK
7,8
TEMP ALARM
9, 10
DIAG MODE
11,12
POWER ON
13,14
PRESS ALARM
15,16
HIGH RANGE
SELECTED
17,18
FLOW ALARM
19,20
RESERVED
SOURCE
21,22
WARNING
23,24
RESERVED
Condition
Gas concentration is exceeding the Hi set
limit.
Gas concentration is exceeding the HiHi set
limit.
True if no alarm conditions exist
True if any temp warning exists
True if in diagnostic mode
True if main power is on
True on low pressure
True if the auto-range function has switched
the analyzer into high range.
True if a flow warning exists
True if the analyzer source intensity is out
of limits.
1-12