Completing the Installation

Completing the Installation
22
Surveyor PDA Plus Detector Hardware Manual
The Surveyor PDA Plus Detector is calibrated at the factory. However, it requires
recalibration after it is installed and before it is used for analysis. The calibration is dependent
on the exact physical alignment of components on the optical bench within the unit. Bumps
and jars, which might occur in shipping and during the installation process, can affect this
alignment.
The detector is calibrated from the data system you are using to control it. Before you can
calibrate the detector, you must add it to the instrument configuration for your data system.
For details on adding the detector to your instrument configuration, see
"Configuration."
For details on calibrating your detector, see
Detector."
Chapter 4, "Xcalibur Diagnostics for the PDA
Chapter 3,