Xcalibur Diagnostics for the PDA Detector

Calibrating the PDA Detector

Performing a Wavelength Calibration

Thermo Scientific
Use the diagnostics program to check the operation of the PDA detector, calibrate the
detector, and view the event log. To perform diagnostics from the Xcalibur data system, your
detector must be connected to the data system computer, configured as part of an instrument,
and turned on. For information on configuring the PDA detector, see
"Configuration."
Contents
Calibrating the PDA Detector
Retrieving, Viewing, Printing, and Clearing the Event Log
Creating a Display Method to View the Light Intensity
Verifying Operational Performance
Controlling the Lamps with Xcalibur
Adjusting the Attenuators
Checking the Firmware Version
To calibrate your PDA detector, perform the following procedures:
Performing a Wavelength Calibration
Performing a Dark Current Calibration
The alignment of the spectrum on the diode array is dependent upon the physical alignment
of various components of the optical bench. The alignment can become offset if the detector
is sharply jolted, for example, in shipping. Such bumps and jars can slightly change the
wavelength of light reaching the photodiode array. Use the automated wavelength calibration
to determine wavelength accuracy and to correct for any misalignment.