Determining the Diode of Maximum Intensity for the UV Range

Thermo Scientific
c. Click Default, and then verify that the following parameters are specified:
Start = 2
End = 511
Step = 1
Bandwidth = 1
Scan rate = 1
d. Click Load To Detector.
To determine the diode of maximum intensity for the UV Range
1. Click the Display tab to open the Display page.
2. Click Start, which is located in the upper right corner of the Display page.
3. Right-click in the top pane, which displays the intensity spectrum, to open the shortcut
menu.
4. From the shortcut menu, choose Axis Setup to open the Axis Properties dialog box (see
Figure
72).
Figure 72. Axis Properties dialog box
5