Configuration

Thermo Scientific
The only manual control on the Surveyor PDA Plus Detector is the On/Off switch below the
left door of the unit. All other instrument control functions are performed from the
chromatography data system.
To control the detector, you must add it to the instrument configuration for your data system.
The only configurable option for the PDA detector is its stack address, which must match its
unit ID. The unit ID is set by adjusting the two rotary switches on the back panel of the
detector. As
Figure 18
shows, the switches are set to 01 at the factory. Do not adjust the
switches unless you are controlling more than one Surveyor PDA Plus Detector from the same
data system computer.
Figure 18. Unit ID set to 01
Contents
Configuring the Surveyor PDA Plus Detector in Xcalibur
Configuring the Surveyor PDA Plus Detector in ChromQuest
Surveyor PDA Plus Detector Hardware Manual
3