5
ChromQuest Diagnostics for the PDA Detector

Verifying Operational Performance

Verifying Operational Performance

Verifying the Performance of the Detector

72
Surveyor PDA Plus Detector Hardware Manual
If you move your PDA detector, replace lamps, or install a new flow cell, the system
performance might change. Verify that the detector is running as desired by performing the
following procedures:
Verifying the Performance of the Detector
Recording the Performance of the Lamps
To verify the proper operation of the detector from ChromQuest
1. Open the online Instrument window for your instrument.
2. Create and download a method that pumps 100% HPLC-grade methanol or
HPLC-grade water at a constant flow rate of 1 mL/min.
3. Check the status of the lamps:
a. On the menu bar, choose Control > Instrument Status to display the Instrument
Status window.
b. Click the Surveyor PDA Plus tab to display the Surveyor PDA Plus Instrument
Status page.
c. Click Diagnostics to open the PDA Plus Diagnostics dialog box.
d. Click the Lamps tab to display the lamps page. Note the status and usage for each
lamp.
e. If they are not already on, turn on the deuterium (D
for 1 hour for both lamps to equilibrate.
4. Perform a dark current calibration, and then perform a wavelength calibration. Use the
wavelength calibration file appropriate for your normal operation. Do not forget to apply
the results by clicking Apply.
5. View an intensity scan of the deuterium (D
a. Click the Lamps tab to display the Lamps page.
b. If the tungsten lamp is not off, turn it off.
c. Click the Control tab to display the Control page.
d. In the Mode area, select the Intensity option to set the units to intensity.
e. Click Default.
f.
In the Discrete Channels area, set Channels A, B, and C to monitor diodes 35, 52,
and 77 respectively (see
) lamp:
2
Figure
67).
) and tungsten (W) lamps. Wait