Body weight [kg]
126 to 150 (277 lbs to
330 lbs)

2.3 Environmental conditions

1C60
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
1C63, 1C64
Allowable environmental conditions
Temperature range: –10 °C to +45 °C (14 °F to 113 °F)
Chemicals/liquids: fresh water, salt water, perspiration, urine, soapsuds, chlorine
water
Moisture: Submersion: max. 1 h in 3 m depth, relative humidity: no restrictions
Solids: Dust, occasional contact with sand
Clean the product after contact with humidity/chemicals/solids, in order to
avoid increased wear and damage (see Page 21).
Unallowable environmental conditions
Solids: Highly hygroscopic particles (e.g. talcum), continuous contact with sand
Chemicals/liquids: Acids, continuous use in liquid media
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative humidity 20 % to
90 %, no mechanical vibrations or impacts
2.4 Service life
This product has been tested by the manufacturer with 2 million load cycles
according to ISO 22675. Depending on the patient's activity level, this cor­
responds to a service life of 2 to 3 years.
Spring stiffness
Prosthetic foot size [cm]
21
22
23
24
25
26
27
28
5
29
30
15

Hide quick links: