Download Print this page

New Functionality For Cpu Firmware Version 3.600 And Above - Thermo Scientific DensityPRO-T Series User Manual

Gamma density measurement systems

Advertisement

Caution: Using this equipment in a manner not specified by
Thermo Scientific may impair the product, leading to
equipment malfunction.
1. New Functionality for CPU Firmware
Version 3.600 and above.
NOTE: Please read the notes below before upgrading from CPU firmware
≤ 3.500 to CPU firmware version 3.600 and above.
With the release of CPU firmware Versions ≥ 3.600 we are adding
functionality permitting users to perform standardization (STD) and
calibrations (CAL) in user selected units. To make use of this functionality
requires that the CPU firmware [9-0700-003], Remote Backplane (RBP)
firmware
[9-0700-6] and EZ CAL II software [9-0700-007] all be at version 3.600 or
newer. Failure to upgrade the CPU Remote Backplane and EZ CAL II
software to version 3.600 or newer will limit you to performing STD and
CAL funtions in g/cc.
2. Density Measurement Standardization and
Calibration in User Selected Primary Measurement Type
The following table provides an overview of which CPU firmware in
combination with listed user interfaces support performing
standardization and calibration in user selected density primary
measurement type.