Download Print this page

Thermo Scientific DCT6088 User Manual

Dedicated transit time flowmeter

Advertisement

Quick Links

DCT6088
Dedicated Transit Time Flowmeter
User Guide
P/N 1-0561-006
Revision J
Part of Thermo Fisher Scientific

Advertisement

loading

Summary of Contents for Thermo Scientific DCT6088

  • Page 1 DCT6088 Dedicated Transit Time Flowmeter User Guide P/N 1-0561-006 Revision J Part of Thermo Fisher Scientific...
  • Page 3 DCT6088 Dedicated Transit Time Flowmeter User Guide P/N 1-0561-006 Revision J...
  • Page 5 ©2017 Thermo Fisher Scientific Inc. All rights reserved. “Microsoft”, “Windows”, and “Excel” are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries. “Dow Corning” is a registered trademark of Dow Corning Corporation. “Krautkramer” is a registered trademark owned by GE Inspection Technologies, Llc, Krautkramer Inc.. “Sil-Glyde”...
  • Page 6 This page intentionally left blank.
  • Page 7: Revision History

    Revised per ECO 4760. 02-2006 Revised per ECO 5077. 05-2007 Revised per ECO 5601. 06-2007 Revised per ECO 5739. 11-2007 Revised per ECO 6016. 06-2011 Revised per ECO 7752. Revised per ECO 9242. 07-2017 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 8 This page intentionally left blank.
  • Page 9: Table Of Contents

    Single Phase 240 Vac Operation ............3-3 Double Phase 240 Vac Operation ............ 3-3 12 to 24 Vdc Operation ..............3-3 The Current Loop................3-4 Relay Terminals .................. 3-5 Communications Terminal Block ............3-6 Transducer Wiring................3-6 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 10 Chapter 7 Maintenance & Troubleshooting ..............7-1 Replacing the Fuse ................7-1 Replacing the Current Loop Module........... 7-2 Replacing Sonic Coupling Compound..........7-3 Software Upgrades ................7-4 General Troubleshooting ..............7-4 Contact Information ................7-5 viii DCT6088 User Guide Thermo Fisher Scientific...
  • Page 11 Monitoring & Downloading Data Logs Using D-Link ........C-1 Purpose ....................C-1 Installing D-Link ................C-1 Establishing Communications with a Flowmeter........ C-1 Monitoring Data Logs................ C-2 Saving & Loading Data Logs ............. C-3 Appendix D Toxic & Hazardous Substances Tables............D-1 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 12 This page intentionally left blank.
  • Page 13: Safety Information & Guidelines

    Failure to correctly perform the instructions and procedures in this manual or other documents pertaining to this system can result in equipment malfunction, equipment damage, and/or injury to personnel. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 14: Warnings, Cautions, & Notes

    ▲ Note Notes emphasize important or essential information or a statement of company policy regarding an operating procedure, practice, condition, etc. ▲ DCT6088 User Guide Thermo Fisher Scientific...
  • Page 15: Chapter 1 Product Overview

    Chapter 1 Product Overview Introduction The Thermo Scientific DCT6088 dedicated transit time flowmeter is a microprocessor-based instrument that measures the flow of clean, homogeneous liquids (liquids without large concentrations of suspended particles or gasses such as air bubbles). The flowmeter is non-invasive, which means that it measures flow from outside the pipe.
  • Page 16: Transit Time Accuracy

    While calibration on a reference flow loop under known conditions is a useful exercise to determine the accuracy potential of an instrument, it is not a guarantee of absolute accuracy on different pipes under field conditions. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 17: Ordering Information

    The table below provides ordering information for the flowmeter. Information Table 1–1. Code Model DCT6088 Thermo Scientific DCT6088 dedicated digital correlation transit time flowmeter Code Power Supply 90 to 132 Vac, 50/60 Hz 190 to 250 Vac, 50/60 Hz 12 to 24 Vdc nominal...
  • Page 18: Specifications

    Transmitters: -40°C to 60°C (-40°F to 140°F) Contact factory for higher temperature range requirements. Keypad 19-key with tactile action Display 2-line x 40-character, alphanumeric, backlit LCD Data logger 30,000 point data logger, programmable in 1-second intervals DCT6088 User Guide Thermo Fisher Scientific...
  • Page 19: External Features

    Product Overview External Features External Features Figure 1–3. External features of the DCT6088 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 20 This page intentionally left blank.
  • Page 21: Chapter 2 System Installation

    1. Screw the four mounting ears to the metal threaded inserts on the back of the enclosure using the 5/16” screws provided in the mounting ears kit. 2. Attach the ears to the wall with standard mounting screws. Figure 2–1. Mounting ears method Thermo Fisher Scientific DCT6088 User Guide...
  • Page 22: Transducer Installation

    Transducer Installation Transducer Installation Site Selection & Prior to installing the transducers, a proper site must be selected to ensure accurate measurement. Examples of site recommendations are illustrated Preparation below. Figure 2–2. Site recommendations DCT6088 User Guide Thermo Fisher Scientific...
  • Page 23 If your application cannot follow these guidelines completely, meaningful flow measurements (with some loss in accuracy and stability) may still be obtained, depending on signal quality. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 24: Spacing & Mounting The Transducers

    Reversing the position of the upstream and downstream transducers or reversing the transducer cable connections to the instrument will result in negative flow readings. ▲ DCT6088 User Guide Thermo Fisher Scientific...
  • Page 25 Adjust the transducers until the gaps on both sides are equal. c. While holding the transducers in place, tighten the clamps sufficiently to prevent the transducers from slipping and to allow proper flowmeter operation. Figure 2–4. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 26 RCV GND terminal. 8. If maximum accuracy at low flow rates is important, calibrate the flowmeter according to “The Calibration Menu” (Chapter 5). The flowmeter is now capable of accurately measuring velocity and flow. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 27: Transducer Mounting Methods

    Like the V method, both transducers are mounted on the same side of the pipe. When configuring the flowmeter, ensure W is the selected mounting method. Figure 2–6. W mounting method Thermo Fisher Scientific DCT6088 User Guide...
  • Page 28: Z Method

    2. Place a transducer at the 3 o’clock position. 3. Trace the shape of the 3 o’clock transducer along its inside edge (opposite the cable connection). Draw a horizontal line at its center. Remove the transducer. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 29 Figure 2–10. Z mounting method, steps 4–6 7. Wrap the paper firmly around the pipe, and mark the intersection point where the fold comes in contact with the rest of the paper. Figure 2–11. Z mounting method, step 7 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 30 UltraScan software or via menu 25. For example, if the software calculates the spacing as 14 inches (356 mm), draw a line 17 inches (432 mm) long. Figure 2–14. Z mounting method, step 10 2-10 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 31 13. Go to the other side of the pipe (9 o’clock position), and mark the pipe at the point where the marked fold and the inside edge of the paper length intersect. Figure 2–17. Z mounting method, step 13 Thermo Fisher Scientific DCT6088 User Guide 2-11...
  • Page 32: Wv And Ww Methods

    For applications with pipe diameters smaller than 2 inches (50 mm), the WV and WW methods are options that allow higher accuracy and stability Methods to be achieved when reasonable signal strength can be obtained (10% or higher). 2-12 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 33: Small Pipe Applications

    2. Draw two lines on each transducer surface with a pencil so that the band defined by the lines is in the middle of the surface. The spacing between the two lines should be as shown in the following table. Thermo Fisher Scientific DCT6088 User Guide 2-13...
  • Page 34 Figure 2–23. 4. Apply coupling compound to the space between the tape strips. The optimum height of the compound layer is approximately half the height of the tape strips. Figure 2–24. 2-14 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 35: Chapter 3 Wiring

    Warning All wiring should be routed through conduit or cable glands to seal the enclosure. Refer to Figure 3–1 for the recommended cable routing. Bonding between conduit connections is not automatic and must be provided as part of the installation. ▲ Thermo Fisher Scientific DCT6088 User Guide...
  • Page 36: Power Connections

    To connect power, remove the sheet metal access cover. Locate the individual power input terminals. They are marked directly on the main board or on a sticker attached to the main board. Connect the power according to one of the following sections. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 37: 120 Vac Operation

    ▲ To connect the power for 12 to 24 Vdc operation: 4. Connect the positive wire to the L1(+) terminal. 5. Connect the negative wire to the L2/N(-) terminal. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 38: The Current Loop

    1. Ensure the loop powering option is correct. Locate the jumpers on the upper right corner of the module. Refer to the figure below to change the jumper settings. Figure 3–2. Jumper settings for current loop configurations DCT6088 User Guide Thermo Fisher Scientific...
  • Page 39: Relay Terminals

    (normally closed). When the relay is energized, the C terminal is shorted to the NO terminal (refer to Figure 3–5). When the relay is at rest, the C terminal is shorted to the NC terminal. Figure 3–5. Relay schematic Thermo Fisher Scientific DCT6088 User Guide...
  • Page 40: Communications Terminal Block

    Refer to the figure below. The symbol is on both pairs of terminals and indicates which terminals should connect to the center wire conductors and which should connect to the coaxial shields. Figure 3–7. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 41: Operating & Configuring The Flowmeter

    1. Press MENU followed by the +/- (plus/minus) key. 2. LCD CONTRAST appears on the display with a bar the indicating current contrast setting. 3. Press the left or right arrows to adjust the contrast. 4. Press ENTER when complete. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 42: Flowmeter Configuration

    ENTER. 4. Complete the process by accessing a Primary display (menus 00 through 04). Note The flowmeter will not use the new parameters until you access a Primary display. ▲ DCT6088 User Guide Thermo Fisher Scientific...
  • Page 43 Transducer Transducer Type Transducer Mounting Transducer Spacing Flow Flow Units Max Flow Range Min Flow Range Damping Low Flow Cutoff Low Signal Cutoff Totalizer Totalizer Units Totalizer Multiplier Net Totalizer Positive Totalizer Negative Totalizer Thermo Fisher Scientific DCT6088 User Guide...
  • Page 44 Current Loop Span Current Loop Calibration Current Loop Test Relays Program Relays View Relays Test Relays Datalog Datalog Interval Diagnostics Signal Strength/Margin Delta Time/Fluid Sound Speed Reynolds#/Profile Factor Current Loop Output Software/Firmware Rev. Level DCT6088 User Guide Thermo Fisher Scientific...
  • Page 45: Using The Arrow Keys To Access Menus

    Figure 4–4. Remaining submenus of the Main menu 3. When the desired submenu is highlighted, press ENTER to display the first menu of the selected submenu. The figure below shows what is displayed if the Flow submenu is selected. Figure 4–5. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 46: Quick Setup Configuration

    Actual measurements (not nominal) must be entered. ▲ a. Select the PIPE OD menu (10). The screen shown in Figure 4–6 is displayed. Enter the value for the pipe OD and press ENTER. Press the down arrow, and select Actual. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 47 3. If there is a liner, continue with this step. Otherwise, skip to step 4. a. Access the Liner Material setup menu (16). The screen shown in Figure 4–10 is shown. Select one of the available options. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 48 (21) and fluid viscosity (22). ▲ Figure 4–12. Fluid Type setup menu Note Refer to “The Fluid Menu” (Chapter 5) for additional items not addressed in the quick setup procedure. ▲ DCT6088 User Guide Thermo Fisher Scientific...
  • Page 49 Figure 4–15. Transducer Spacing menu 7. Enter the Flow submenu, and access the Flow Units setup menu (30). Select the flow rate units. a. Select one of the available volumetric units. Figure 4–16. Flow Units setup menu Thermo Fisher Scientific DCT6088 User Guide...
  • Page 50 Note If any of the above setup parameters are changed, the flowmeter stops measuring flow until the new value is entered and a Primary Display is accessed to accept the new value. ▲ 4-10 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 51: Primary Displays & Menus

    The Flow/Positive Totalizer primary display (02) indicates the flow rate and the totalized flow in the positive flow direction. If the positive totalizer Totalizer is not currently enabled, the last net totalized value is displayed. Figure 5–3. Flow/Positive Totalizer primary display Thermo Fisher Scientific DCT6088 User Guide...
  • Page 52: Flow/Negative Totalizer

    Figure 5–7. Pipe Sound Speed setup menu DCT6088 User Guide Thermo Fisher Scientific...
  • Page 53: The Liner Menu

    Figure 5–10. Liner Inside Roughness setup menu Thermo Fisher Scientific DCT6088 User Guide...
  • Page 54: The Fluid Menu

    Figure 5–12. Fluid Viscosity setup menu DCT6088 User Guide Thermo Fisher Scientific...
  • Page 55: The Flow Menu

    If this is the case, increase the damping coefficient just enough to reduce the fluctuation to an acceptable degree. Figure 5–15. Damping setup menu Thermo Fisher Scientific DCT6088 User Guide...
  • Page 56 To set the low signal cutoff, access the Low Signal Cutoff setup menu (35). Enter the low signal cutoff and press ENTER. Figure 5–18. Low Signal Cutoff setup menu DCT6088 User Guide Thermo Fisher Scientific...
  • Page 57: The Total Menu

    For example, 700 liters can be displayed as ‘700’ if the selected multiplier value is X1, or it can be displayed as ‘7’ if the selected multiplier value is X100. Figure 5–21. Totalizer Multiplier setup menu Thermo Fisher Scientific DCT6088 User Guide...
  • Page 58 It is not affected by flow in the opposite direction. Figure 5–24. Negative Totalizer setup menu Totalizer Reset menu (41) ● Use this menu to reset one or all of the totalizers. Figure 5–25. Totalizer Reset menu DCT6088 User Guide Thermo Fisher Scientific...
  • Page 59: Options Menu

    To access a different site, press Enter, scroll to the desired site, and press Enter again. Note To avoid overwriting stored parameters and losing old data, ensure that the desired site is active prior to entering the new set of parameters. ▲ Thermo Fisher Scientific DCT6088 User Guide...
  • Page 60 PC. Communication is established using the Thermo Scientific UltraScan configuration software (refer to the UltraScan user guide) or the Thermo Scientific D-Link interface utility (refer to Appendix C). To configure the RS232 port, access the RS232 Configuration setup menu and select ULTRASCAN or D-Link.
  • Page 61 To change or disable the scale factor password, access the Change Scale Factor Password setup menu. Enter the new scale factor password and press Enter. Figure 5–33. Change Scale Factor Password setup menu Thermo Fisher Scientific DCT6088 User Guide 5-11...
  • Page 62 Unit ID (identification) Number setup menu (49) ● This number is set by the operator to identify the specific instrument or site. Any whole number between 1 and 60,000 may be entered. Figure 5–36. Unit ID setup menu 5-12 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 63: The Calibration Menu

    Prior to performing a zero set calibration, verify the following: Transducers are connected to the pipe. ● Instrument is reading flow. ● Low flow cutoff is disabled to allow verification of calibration. ● Thermo Fisher Scientific DCT6088 User Guide 5-13...
  • Page 64 Figure 5–39. Zero set calibration successful Figure 5–40. Zero set calibration failed 4. If Figure 5–40 is displayed, press any key and repeat steps 1 through 3 when the flow rate is within required limits. 5-14 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 65     11. Access the Zero Set menu (51) and select Manual. 12. Enter the zero point. If necessary, you can apply a negative offset by pressing +/- (plus/minus key). Thermo Fisher Scientific DCT6088 User Guide 5-15...
  • Page 66: Scale Factor Calibration

    To enter a new scale factor, follow the steps below. 1. Access the Scale Factor setup menu (52). Figure 5–41. 2. Enter the new scale factor and press Enter. 5-16 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 67: Sound Speed Compensation

    2. Reconfigure the flowmeter, and access the Date and Time setup menu (54). Note Performing a Master Erase and reconfiguring the flowmeter prior to changing the date or time prevents possible data corruption. ▲ Thermo Fisher Scientific DCT6088 User Guide 5-17...
  • Page 68: The 4 -20 Ma Menu

    1. Connect a milliammeter to the input (IN) and output (OUT) terminals of the current loop module. 2. Access the Current Loop Calibration menu (58). Figure 5–47. 4 mA Calibrate screen 5-18 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 69 If the values do not match, repeat steps 2 through 10. 11. Set the current loop span by accessing the Current Loop Span menu (57). The current loop span is the span of flow versus current. Thermo Fisher Scientific DCT6088 User Guide 5-19...
  • Page 70: The Relays Menus

    Programming To program a relay, follow the steps below. 1. Access the Program Relays menu (71). Figure 5–52. Program Relays menu 5-20 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 71 NET TOTAL PULSE : Relay cycles once each time the net ● totalizer increments. POS TOTAL PULSE : Relay cycles once each time the positive ● totalizer increments. NEG TOTAL PULSE : Relay cycles once each time the negative ● totalizer increments. Thermo Fisher Scientific DCT6088 User Guide 5-21...
  • Page 72 Enter the value for the relay ON condition. Press Enter. Figure 5–54. Relay On Condition screen Note The flow units used for the relays are the same as the flow units selected for measuring flow in the Flow Units menu (30). ▲ 5-22 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 73 6. Press the down arrow, and the Relay Off Condition Value screen is displayed. Enter the value for the relay OFF condition. Press Enter. Figure 5–56. Relay Off Condition Value screen 7. Repeat this procedure for each relay you want to program. Thermo Fisher Scientific DCT6088 User Guide 5-23...
  • Page 74: Viewing Relays

    3. Repeat steps 1 through 2 for each relay you want to test. 4. Once you have completed testing the desired relays, exit the relay test function by pressing Erase or by directly accessing another menu. 5-24 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 75: The Data Log Menu

    Any power interruptions cause zeroes to be written in the data log stream. This feature is useful to determine if an interruption in the time sequence has occurred or if the instrument has been moved or tampered with while logging data. Thermo Fisher Scientific DCT6088 User Guide 5-25...
  • Page 76: The Diagnostics Menu

    The flow profile factor is calculated by the flowmeter and used to determine the effect of the flow profile on mean measured fluid velocity. Figure 5–64. Reynolds Number/Profile Factor display 5-26 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 77 Figure 5–65. Current Loop Output display Software/Firmware Revision Level display (94) ● Displays the software (SOFT VERS.) and firmware (FPGA VERS.) versions installed in the flowmeter. Figure 5–66. Software/Firmware Revision Level display Thermo Fisher Scientific DCT6088 User Guide 5-27...
  • Page 78 This page intentionally left blank.
  • Page 79: Emergency Override & Master Erase

    1. Turn the flowmeter off and back on. 2. When the message INITIALIZING... is displayed, press ERASE within three seconds. 3. The Master Erase screen is displayed. To continue, press the 5 key within three seconds. Figure 6–1. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 80 4. At the prompt, press the . (decimal) key within three seconds to continue. Figure 6–2. 5. If the master erase function is completed, the message shown below is displayed. Figure 6–3. 6. Enter all configuration data. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 81: Chapter 7 Maintenance & Troubleshooting

    4. Using a small, flat-bladed screwdriver, remove the fuse and replace it with another fuse of the same rating. 5. Reconnect power to the flowmeter and verify that the unit is operating properly, e.g. the replacement fuse does not blow. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 82: Replacing The Current Loop Module

    It is possible to incorrectly align the module by one pin and force the holes to line up, causing damage to the module. ▲ See the figure below. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 83: Replacing Sonic Coupling Compound

    Follow the steps below to replace the coupling compound. 1. Remove the transducers from the pipe. 2. Clean the old compound from the transducers and the pipe. 3. Apply a wide bead of compound lengthwise down the center of each transducer face. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 84: Software Upgrades

    1. Verify the flowmeter is properly installed and the installation site is suitable. 2. Verify the flowmeter is properly configured. 3. Perform a master erase (Chapter 7). 4. Contact the installation contractor or representative through whom the flowmeter was purchased. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 85: Contact Information

    Ion Path, Road Three Senapati Bapat Road Winsford, Cheshire Pune 411 016 CW7 3GA Maharashtra, INDIA UNITED KINGDOM +91 (20) 6626 7000 +44 (0) 1606 548700 +91 (20) 6626 7001 fax +44 (0) 1606 548711 fax www.thermoscientific.com Thermo Fisher Scientific DCT6088 User Guide...
  • Page 86: Warranty

    Maintenance & Troubleshooting Warranty Warranty Thermo Scientific products are warranted to be free from defects in material and workmanship at the time of shipment and for one year thereafter. Any claimed defects in Thermo Scientific products must be reported within the warranty period. Thermo Fisher Scientific shall have the right to inspect such products at Buyer’s plant or to require Buyer to...
  • Page 87: Hazardous Area Installation

    Instrument should be installed in accordance with National Electrical Code (NEC) Article 500, including use of explosion proof seals for the wiring connections to the enclosure. The flowmeter must be installed and wired in accordance with the specified installation drawing. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 88 ▲ Figure 8–1. North American IS (Div. 1) hazardous area installation (refer to drawing 22493-0003) DCT6088 User Guide Thermo Fisher Scientific...
  • Page 89 Hazardous Area Installation North American Certification Figure 8–2. North American non-incendive (Div. 2) hazardous area installation (refer to drawing 22493-0001) Thermo Fisher Scientific DCT6088 User Guide...
  • Page 90: European Certification

    Instrument should be installed in accordance with required codes, including use of explosion proof seals for the wiring connections to the enclosure. The flowmeter must be installed and wired in accordance with the specified installation drawing. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 91 Hazardous Area Installation European Certification Figure 8–3. European hazardous area installation (refer to drawing 22493-0005) Thermo Fisher Scientific DCT6088 User Guide...
  • Page 92: North American Hazardous Area Installation Definitions

    65 gal/min at a 10-ft distance for 5 minutes; additional corrosion-resistant characteristics, having no exposed metal surfaces NEMA 7: Explosion proof enclosures for indoor hazardous locations ● (Class I, Groups A, B, C, D) DCT6088 User Guide Thermo Fisher Scientific...
  • Page 93: European Hazardous Area Installation Definitions

    Non user-replaceable fuse for all versions: 5 A, 250 V, quick acting Output isolation 5 kVac surge isolation when powered from an external source; not to be connected to continuous voltages in excess of 50 Vac with respect to ground (earth) Thermo Fisher Scientific DCT6088 User Guide...
  • Page 94 This page intentionally left blank.
  • Page 95: Appendix A Pipe Schedules

    The actual pipe ID may vary from the dimension listed in the tables by as much as 25% of the pipe minimum wall thickness. The accuracy of flow rate measurement is enhanced if the pipe ID is actually measured. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 96 Pipe Schedules Table A–1. Steel, stainless steel, and PVC pipe standard schedules DCT6088 User Guide Thermo Fisher Scientific...
  • Page 97 Pipe Schedules Table A–2. Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide...
  • Page 98 Pipe Schedules Table A–3. Ductile iron pipe standard classes DCT6088 User Guide Thermo Fisher Scientific...
  • Page 99: Appendix B Fluid Properties

    2.85 Conc.-glacial 1.34 Acetic anhydride 1384 4540 0.88 Acetone CH COCH 1190 3903 0.41 Acetylene tetrabromide 1007 3303 Acetylene tetrachloride 1155 3788 Alcohol allyl 1.60 0.90 butyl-n 3.64 1.17 ethyl (grain) C 1.52 37.8 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 100 647 – 1295 RS-1, MS-1, SS-1 33 – 216 37.8 19 – 75 Asphalt emulsions Fed #1 215 – 1510 37.8 75 – 367 Fed #2, V, VI 33 – 216 37.8 19 – 75 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 101 11.6 Bromine 0.34 Bromobenzene 1074 3523 Bromoform 2978 Butane-n -1.1 0.52 0.35 Butyl acetate 1172 3844 n-Butyl alcohol 1257.7 4125 iso-Butyl bromide -104 1450 -155 4756 Butyric acid n 1.61 2.3cp Cadmium 2150 7052 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 102 Coconut oil 37.8 29.8 – 31.6 54.4 14.7 – 15.7 Cod oil 37.8 32.1 54.4 19.4 Corn oil 54.4 28.7 Corn starch solutions 22 Baume 21.1 32.1 37.8 27.5 24 Baume 21.1 129.8 37.8 95.2 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 103 1 – 3.97 37.8 6 – 11.75 54.4 3.97 – 6.78 37.8 29.8 max 54.4 13.1 max 86.6 max 71.1 35.2 max Diethyl ether 0.32 Diethylene glycol 21.1 1533 5028 Diethylene glycol monoethyl 1296 4251 ether Thermo Fisher Scientific DCT6088 User Guide...
  • Page 104 1616 5300 Ethylene glycol monoethyl 1279 4195 ether Ethylene glycol monomethyl 1339 4392 ether Formaldehyde 1587 5205 Formamide 1610 5281 Formic acid 1299 4261 1.04 Conc. 1.48 Freon 21.1 0.21 21.1 0.27 21.1 1.45 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 105 37.8 0.71 15.6 0.64 15.6 0.46 37.8 0.40 Glycerine 1923 6307 100% 20.3 37.8 50% water 5.29 1.85cp Glucose 37.8 7.7M – 22M 65.6 880 – 2420 Guaicol 1252 4107 Helium -268.8 179.8 -452 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 106 93.3 Machine lubricants 37.8 23 – 34 54.4 13 – 18 37.8 34 – 72 54.4 18 – 25 37.8 72 – 83 54.4 25 -39 37.8 75 – 119 54.4 39 – 55 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 107 Menthol 1271 4169 Merck 20.2 1482.3 4862 Mercury 1454 4769 21.1 0.118 37.8 0.11 Methanol 1118 3667 Methyl acetate 0.44 1131 3710 Methyl alcohol 1121.2 3678 Methyl bromide 2968 Methylene chloride 23.5 1064 3490 Thermo Fisher Scientific DCT6088 User Guide...
  • Page 108 1625 5330 Oil of camphor 1390 4559 Oleic acid 1442 4730 Olive oil 21.7 1440 4723 37.8 43.2 54.4 24.1 Oxygen -182.9 -297 2991 Palm oil 37.8 47.8 54.4 26.4 Paraldehyde 1197 3926 B-10 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 109 54.4 Rosin oil 37.8 324.7 54.4 129.9 Rosin (wood) 37.8 216 – 11M 93.3 108 – 4400 Rubidium 1260 4133 Sesame seed oil 37.8 39.6 54.4 Silicon tetrachloride 766.2 2513 Sodium 2500 8200 Thermo Fisher Scientific DCT6088 User Guide B-11...
  • Page 110 37.8 3200cp 82.2 160cp Sucrose 60 Brix 21.1 49.7 37.8 18.7 64 Brix 21.1 49.7 37.8 31.6 68 Brix 21.1 216.4 37.8 59.5 72 Brix 21.1 37.8 138.6 74 Brix 21.1 1210 37.8 B-12 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 111 4.4M – 13.2M 53.7 – 86.6 RT-12 25M – 75M 108 – 173 Tetralin 1484 4868 Tin (molten) 2470 8102 Toluene 0.68 1275 4182 o-Toluidine 22.5 1669 5474 l-Tridecene 1313 4307 Trielhylene glycol 21.1 Thermo Fisher Scientific DCT6088 User Guide B-13...
  • Page 112 Fresh 15.6 1.13 54.4 0.55 1.15 Water (sea) 1507.4 4944 (Surface, 3.5% salinity) Whale oil 37.8 35 – 39.6 54.4 19.9 – 23.4 Xylene hexafluoride 2883 o-Xylene 0.93 1352 4435 Zinc 2700 8856 B-14 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 113: Clean Water Sound Speed Versus Temperature

    1481.968 4909.1 1496.294 4949.9 1508.73 4985 1519.428 5014.7 1528.481 5039.4 1536.009 5059.6 1542.166 5075.4 1546.982 5087.2 1550.579 5095.3 1553.047 5099.7 1554.389 5100.8 1554.724 5098.7 1554.084 5093.6 1552.529 5085.6 1550.091 5074.8 1546.799 5068.4 1544.848 Thermo Fisher Scientific DCT6088 User Guide B-15...
  • Page 114: Relationship Between Specific Gravity, Viscosity, & Sound Velocity For Petroleum Products

    Figure B–1. Graphical representation of Table B–2 Relationship The graph below illustrates the approximate relationship between specific gravity, viscosity and sound velocities for aliphatic hydrocarbons between (petroleum products). Specific Gravity, Viscosity, & Sound Velocity for Petroleum Products B-16 DCT6088 User Guide Thermo Fisher Scientific...
  • Page 115: Appendix C Monitoring & Downloading Data Logs Using D-Link

    Monitoring & Downloading Data Logs Using D-Link Purpose The Thermo Scientific D-Link utility is a PC-based user interface software that can communicate with the Thermo Scientific DCT6088 and DCT7088 flowmeters. This appendix provides instructions on how to install the software and use it to monitor and download data log information.
  • Page 116: Monitoring Data Logs

    This enables you to monitor the flow data without stopping the log. Refer to “The Data Log Menu” (Chapter 5) for instructions on setting up the data logger. DCT6088 User Guide Thermo Fisher Scientific...
  • Page 117: Saving & Loading Data Logs

    Once the data log is displayed, you can save it. Go to File > Save Log. The file can then be opened in data processing software such as Microsoft® Loading Data Excel®. To load a previously saved log file, go to File > Load Log. Logs Figure C–4. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 118 This page intentionally left blank.
  • Page 119: Appendix D Toxic & Hazardous Substances Tables

    Appendix D Toxic & Hazardous Substances Tables The English and Chinese versions of the Toxic and Hazardous Substances tables are shown below. Thermo Fisher Scientific DCT6088 User Guide...
  • Page 120 This page intentionally left blank.
  • Page 121 Thermo Fisher Scientific 81 Wyman Street P.O. Box 9046 Waltham, Massachusetts 02454-9046 United States www.thermofisher.com...