Thermo Scientific
Calculated Values
L: Pipe ID (cm) – Inside diameter of the pipe converted from user selected
units to centimeters.
(Dc): Density of the carrier material – For a single-phase material Dc = 0
(g/cm3).
(µc): Attenuation of carrier material – For a single-phase material µc = 0
(cm2/g).
(Ds): Density of the solid material – For a single-phase material Ds = 1
(g/cm3).
(Dcal1): Calibration point 1 Density – Density for calibration point 1 (Dc1u)
converted to grams per cubic centimeter (g/cm3).
(Dcal2): Calibration point 2 Density – Density for calibration point 2 (Dc2u)
converted to grams per cubic centimeter (g/cm3).
D: Density of process material – The density of the process material in g/cm3
calculated using
In addition to Measurement #1, three additional measurements, Measurements #2 –
#4 are available to display additional parameters measured or calculated by the gauge.
These additional measurements are also capable of displaying the values from the
analog inputs.
Density – Displays the density in units from the list in
Bulk Density – Bulk density is the same as density for a single-phase material type.
Temperature – Displays the value of an analog input or the RTD input in the user
selected units.
Degree C (Deg C)
Degree F (Deg F)
Flow Velocity – Displays the velocity of the input configured as a flow type in the
user-selected units.
Feet per second (ft/sec)
Meters per second (m/sec)
Equation
4.
Density Calculation
Single-Phase Material
Units for Measurement