5.4 Check Wiring and Close Access Cover
Double-check all of your electrical connections. Pay attention to polarity of all inputs and outputs. Additionally,
insure that all terminals are clamping onto the bare conductor and not on its insulation. Ensure that wires will not be
trapped or pinched when front cover is replaced and secured. Ensure that excess insulation is not removed from the
wires, as this can lead to poor connections or faulty operation.
Close the main access cover and secure the 2 screws.
6. System Configuration
6.1 Overview
Once all electrical connections have been made, your MPC VECTOR is ready for initial setup, calibration,
and operation. All setup occurs using the handheld remote user interface. The following sections detail the
procedures required to complete the MPC VECTOR configuration.
WHEN POWER IS SUPPLIED TO THE UNIT, LINE VOLTAGE IS PRESENT WITHIN THE MPC VECTOR
ENCLOSURE EVEN WHEN THE PUMP MOTOR IS OFF.
During Start-up, it is necessary to run the pump motor. This will cause fluid to discharge from the pump.
You are responsible for safely diverting flow from the pump during start-up and calibration.
6.2 Critical System Configuration Steps
For proper operation of the MPC VECTOR the following parameters and calibrations must be entered and/or
verified (recommended order of setup shown):
Step Number
Figure 5-12– Sample Analog Output Connections
Parameter(s)
27
Reference IOM
Section