II 3058-4
CHAPTER 9
CHECKING THE OPERATION OF THE PROCESSOR AND
INSTALLING THE CHEMICALS
1.
To allow correct operation of the LEVEL SENSOR, add 500 ml (8 fl oz) of:
•
developer to the DEVELOPER TANK
•
fixer to the FIXER TANK
2.
Slowly fill the DEVELOPER and FIXER TANKS with water to the overflow limit.
Warning
Dangerous Voltage
3.
Energise the main power.
(a)Unlock the MAIN POWER DISCONNECT SWITCH and move it to the "1" position.
(b) Move the CIRCUIT BREAKER CB1 to the "1" position.
4.
Use the LAP TOP COMPUTER diagnostics to check the operation of the following
components:
•
DRYER BLOWER
•
REPLENISHMENT PUMPS
•
RECIRCULATION PUMP - Operate the PUMP for a minimum of 10 seconds to
remove any air in the system.
•
WASH WATER SOLENOID
•
DEVELOPER SOLENOID
•
COVER INTERLOCK SWITCH
•
DEVELOPER and FIXER SENSORS
5.
Check:
•
that the water drains smoothly
•
that the plumbing system has no leakage
6.
Deenergise the main power.
KODAK AG Stuttgart
INSTALLATION INSTRUCTIONS
9-1
8/1998