2.5 Software loading

The program memory on the CPU board is loaded with monitor software at the factory. The
software is used for running all the functions that are integrated into the PC board. For service
and upgrade procedures, the CPU board is fitted with a PCMCIA card drive through which new
software can be loaded.

2.6 Parameter modules

Module
Analog
electronics
Peripheral
drivers
Figure 4
The detailed structure of a parameter module depends on the specific needs for each
individual parameter. However, some common parts are used in the parameter modules. The
electronics inside the module is usually divided into isolated (floating) and non-isolated
sections. Typically, the non-isolated section consists of buffers to interface the parameter
module to the module bus while the rest of the electronics is located in the isolated section. The
isolated section includes the microcontroller together with memory components, the front-end
analog electronics (amplifiers, etc.) and sensor drivers.
keys
A/D
converter
General structure of parameter modules with patient isolation
+5 V
+12 V
Isolation
transformer
CPU
RAM
EEPROM
isolation
System description
+13...16 V
VMOD
Data
Opto
RS485
drivers
Document no. M1144951-004