Memcards; Status - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors

1.3 MemCards

1.3.1 Status

Module Present indicates whether the MemCard
option is installed into the monitor. Possible values
are YES and NO.
Module Active indicates whether the MemCard
services are available. Possible values are YES and
NO.
ROM indicates the status of the ROM memory of the
module. Possible values are OK and ERR.
RAM indicates the status of the RAM memory of the
module. Possible values are OK and ERR.
PCMCIA indicates the status of the PCMCIA controller
of the module. Possible values are OK and ERR.
EEPROM indicates the status of the EEPROM memory
of the module. Possible values are OK and ERR.
SLOT1 and SLOT2 indicate the left hand slot and the right hand slot, respectively.
Card type
indicates whether the card is MENU or DATA card. If a duplicated card is inserted, type DUPL.
File system
indicates the type of the memory card in use. The only supported file system is ATA. If a
memory card using another file system is used, the message 'UNKNOWN' is shown. If the
card is poorly attached, the message 'LOOSE' is shown.
Card size
indicates the total amount of disk space in the card in kilobytes.
Card used
indicates the total amount of used disk space in the card in kilobytes.
Card full
indicates whether all the disk space in the card is used. Possible values are YES and NO.
Card empty
indicates the lack of menu files in the MENU card or no files in the DATA card. Possible values
are YES and NO.
Read error
indicates whether the reading from the card has failed. Possible values are YES and NO.
Write error
indicates whether the writing to the card has failed. Possible values are YES and NO.
All values can be '- - -' to indicate 'No data available'.
10
.
Document no
M1144955-001

Advertisement