Software License, L-Cane06, L-Cane06A, L-Cicu06, L-Cicu06A; Display Controller Section; Display Memory; Display Controller Resolution - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2.3 Software License, L-CANE06, L-CANE06A, L-CICU06, L-
CICU06A
L-CICU06/L-CICU06A software is developed for critical care monitoring and L-CANE06/L-
CANE06A is for anesthesia monitoring. Anesthesia record keeping capability is included in the
anesthesia monitor software L-CANE06/L-CANE06A as standard. The software L-CANE06A and
L-CICU06A include extended arrhythmia analysis.
L-CANE06/L-CANE06A/L-CICU06/L-CICU06A is a license software. The license agreement that
is delivered with the software should be archived in a secure location. A relevant license
number may have to be referred when contacting GE Healthcare service/support. The license
number is needed also for future software upgrades.
Each of the above listed monitor software is loaded on a Central Processing Board at the
factory, i.e. they are not available separately.
The L-xxx06(A)S software is used for service purposes only and is stored on a separate
Software Card (PCMCIA, 8 MB). See section "Downloading the software on empty Central
Processing Board".
The U-xxx06(A) software is used for upgrade purposes and is stored on a separate Software
Card (PCMCIA, 8MB). See section "Downloading the software on empty Central Processing
Board".
CAUTION
The software card is not write protected. For safety reasons do not use
software cards for any other purpose, or on any other platforms than they are
designed for.

2.4 Display controller section

The display controller is integrated to the Central Processing Board. The processor on the CPU
section transmits graphics information through the CPU bus to the display memory on the
Display Controller section.
The Display Controller section supports SVGA resolution. The resolution is automatically
selected for the LCD display in use. The resolution is set to 800 x 600.

2.4.1 Display memory

The Display memory contains digital display data. The size of the memory is 1 Mbyte making
an 800 x 600 display resolution with 256 colors possible. The memory consists of a 1Mbyte
DRAM.

2.4.2 Display controller resolution

The resolution of the display controller is set to SVGA, i.e. 800x600 resolution.
8
.
Document no
M1144956-003

Advertisement

Table of Contents
loading

Table of Contents