4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit
It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact
Anesthesia Monitor via the serial interface port. The data link is bi-directional. The parameters
transferred to the Anesthesia Delivery Unit are summarized in
transferred to the S/5 Compact Anesthesia Monitor are summarized in
transferred continuously once a minute or during the system check procedure and appear in
the trend pages or in the anesthesia record. The transfer of events requires record-keeping
software.
4.3.1 Interconnection
Use the ADU-AS/3 Monitor Cable.
4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit
1.
Press the
2.
Select Install password 10.
3.
Select Interfacing and choose S/5 AM.
4.
Press the
4.3.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor
No settings are required on the S/5 Compact Anesthesia Monitor.
Table 7
Waveforms
CO
2
Setup
key.
Normal Screen
key.
Parameters transferred from S/5 Compact Anesthesia Monitor to S/5
Anesthesia Delivery Unit
Numerics
Et & Fi CO
%
2
Et & Fi O
%
2
O
% difference
2
(I-E)
Et & Fi N
O%
2
Et & Fi AA%
Vol Exp MV
Vol Exp TV
Interfacing
Table 7
and the events
Table
8. The events are