27. Monitor temperature
Check that the shown Monitor Temperature value is within 20 - 40.
28. Battery operation
Disconnect the power cord (without switching the monitor to standby).
Check that the monitoring continues normally. The battery indicator should appear in the
upper right hand corner of the screen.
29. Reconnecting the power cord
Reconnect the power cord.
Check that the monitoring continues without problems. The battery indicator should be
replaced by the text CHARGE.
Recorder Unit
30. Cover open recognition
Open the recorder unit cover. Check that the message 'Recorder: Cover open' appears on
the screen, then close the cover.
31. Recording
Press the key
waveforms. Press the key
NOTE: InvBP waveforms are not recorded without a patient simulator.
NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one
side of the paper is printable.
32. Paper speed
Press the key
that the length of the recorded waveform scale is 18.7 cm (± 1.5 cm).
Change the paper speed setting to 1 mm/s:
Record/Print
Record Waveforms
Press the key
of the scale is now 3.0 cm (± 0.5 cm).
Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
Record Wave
and check that the module starts recording the selected
Stop
on the module to stop recording.
Record Wave
again and this time wait until the recording stops. Check
Paper Speed --> 1 mm/s
RECORD WAVE
and wait until the recording stops. Check that the length
Document no. M1144956-003
27