Temperature measurement
3.
Check the temperature channels with a patient simulator.
Check that the temperature measurement information is shown on the monitor screen as
configured on the simulator.
5.3.11 Dual pressure Module, E-PP/ M-PP
1.
Check the function of the front panel
2.
Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
5.3.12 Cardiac Output Modules, E-COP/ M-COP, E-COPSv/ M-COPSv
Invasive blood pressure measurement
1.
Check the function of the front panel
2.
Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
Cardiac Output measurement
3.
Check the function of the front panel
5.3.13 NIBP module, M-NIBP
1.
Check the function of the front panel membrane keys.
2.
Attach an adult NIBP cuff onto your arm and perform one NIBP measurement.
Check that the module identifies the cuff, i.e. the text 'Adult ' appears on the NIBP digit
field for a short time.
Check that the module gives a reasonable measured result.
5.3.14 Nellcor Compatible Saturation module, E-NSATX/ E-NSAT/ M-NSAT
1.
Connect a Nellcor SpO2 finger probe to the module. Check that the message 'Pulse
search ' is shown. Check also that the shown message changes to 'Check probe ' within
30 seconds.
Zero P5
Zero P6
and
keys.
Zero P4
key.
Start C.O.
key.
Functional check
Document no. M1144951-004
61